Class II (Laminar Flow) Biosafety Cabinetry
Cabinet Style A = Bench Unit With Base or Adjustable Legs Provided B = Bench Unit Without Base or Adjustable Legs Provided C = Console |
Window Type H = Hinged S = Sliding F = Fixed |
Bench Height NA = Not Applicable NP = Not Provided |
CBV = Concurrent Balance Value
Due to a change in nomenclature in NSF/ANSI 49 - 2002 “Class II (Laminar Flow) Biosafety Cabinetry,” Class II, Type A cabinets are now referred to as Class II, Type A1 and Class II, Types B3 and A/B3 cabinets as Class II, Type A2. Class II, Types B1 and B2 cabinets remain unchanged.
Biosafety cabinet models Listed under NSF/ANSI 49 are approved for use with a power supply of 115V/60Hz, unless otherwise noted. If biosafety cabinets have been approved for use with additional power supplies, the power supply will be indicated in the Official Listing.
Labconco Corporation | ![]() |
8845 Prospect | |
Kansas City, MO 64132 | |
United States | |
816-333-8811 X771 | |
Visit this company's website |
Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
30231xxx1 Logic+[1] [2] | A2 | A | 100 - 110 | 50 - 60 | N/A | 3 | 10S | 36 | I.V. Pole U.V. Light Canopy Connection |
30238xxx1 Logic+[1] [3] | A2 | A | 100 - 110 | 50 - 60 | N/A | 3 | 8S | 36 | I.V. Pole U.V. Light Canopy Connection |
30241xxx1 Logic+[1] [3] | A2 | A | 100 - 110 | 50 - 60 | N/A | 4 | 10S | 36 | Canopy Connection I.V. Pole U.V. Light |
30242xxx1 Logic+[3] [4] | A2 | A | 100 - 110 | 50 - 60 | N/A | 4 | 12S | 36 | Canopy Connection I.V. Pole U.V. Light |
30248xxx1 Logic+[1] [3] | A2 | A | 100 - 110 | 50 - 60 | N/A | 4 | 8S | 36 | Canopy Connection I.V. Pole U.V. Light |
30251xxx1 Logic+[1] [3] | A2 | A | 100 - 110 | 50 - 60 | N/A | 5 | 10S | 36 | I.V. Pole U.V. Light Canopy Connection |
30252xxx1 Logic+[3] [5] | A2 | A | 100 - 110 | 55 - 65 | N/A | 5 | 12S | 36 | Canopy Connection I.V. Pole U.V. Light |
30258xxx1 Logic+[1] [3] | A2 | A | 100 - 110 | 50 - 60 | N/A | 5 | 8S | 36 | Canopy Connection I.V. Pole U.V. Light |
30261xxx1 Logic+[1] [3] | A2 | A | 100 - 110 | 50 - 60 | N/A | 6 | 10S | 36 | Canopy Connection I.V. Pole U.V. Light |
30262xxx1 Logic+[3] [4] | A2 | A | 100 - 110 | 50 - 60 | N/A | 6 | 12S | 36 | Canopy Connection I.V. Pole U.V. Light |
30268xxx1 Logic+[1] [3] | A2 | A | 100 - 110 | 50 - 60 | N/A | 6 | 8S | 36 | Canopy Connection I.V. Pole U.V. Light |
30348xxx1 Logic+[1] [3] | B2 | A | 100 - 110 | 50 - 60 | 852 @ 1.8 | 4 | 8S | 36 | Bag-In/ Bag-Out I.V. Pole U.V. Light |
30368xxx1 Logic+[1] [3] | B2 | A | 100 - 110 | 50 - 60 | 1265 @ 2.2 | 6 | 8S | 36 | I.V. Pole U.V. Light Bag-In/ Bag-Out |
30448xxx1 Axiom[3] [6] | C1 | A | 100 - 110 | 60 - 70 | N/A | 4 | 8S | 36 | I.V. Pole U.V. Light Bag-In/ Bag-Out |
30441xxx1 Axiom[3] [6] | C1 | A | 100 - 110 | 60 - 70 | N/A | 4 | 10S | 36 | I.V. Pole U.V. Light Bag-In/ Bag-Out |
30468xxx1 Axiom[3] [4] | C1 | A | 100 - 110 | 50 - 60 | N/A | 6 | 8S | 36 | I.V. Pole U.V. Light Bag-In/ Bag-Out |
30461xxx1 Axiom[3] [6] | C1 | A | 100 - 110 | 60 - 70 | N/A | 6 | 10S | 36 | I.V. Pole U.V. Light Bag-In/ Bag-Out |
31231xxx1 PuriCare[1] [2] | A2 | A | 100 - 110 | 50 - 60 | N/A | 3 | 10S | 36 | I.V. Pole U.V. Light Canopy Connection |
31241xxx1 PuriCare[1] [3] | A2 | A | 100 - 110 | 50 - 60 | N/A | 4 | 10S | 36 | Canopy Connection I.V. Pole U.V. Light |
31242xxx1 PuriCare[3] [4] | A2 | A | 100 - 110 | 50 - 60 | N/A | 4 | 12S | 36 | I.V. Pole U.V. Light Canopy Connection |
31251xxx1 PuriCare[1] [3] | A2 | A | 100 - 110 | 50 - 60 | N/A | 5 | 10S | 36 | Canopy Connection I.V. Pole U.V. Light |
31252xxx1 PuriCare[3] [7] | A2 | A | 100 - 110 | 55 - 65 | N/A | 5 | 12S | 36 | Canopy Connection I.V. Pole U.V. Light |
31261xxx1 PuriCare[1] [3] | A2 | A | 100 - 110 | 50 - 60 | N/A | 6 | 10S | 36 | Canopy Connection I.V. Pole U.V. Light |
31262xxx1 PuriCare[3] [4] | A2 | A | 100 - 110 | 50 - 60 | N/A | 6 | 12S | 36 | Canopy Connection I.V. Pole U.V. Light |
[1] Beginning with serial number 190500000. Inflow nominal set point of 105 fpm was established with a direct airflow reading instrument. This nominal set point was confirmed using the manufacturer's recommended alternate method with thermal anemometer in a constricted access opening, using the manufacturer-supplied probe holder and sash stop (consult manufacturer for appropriate correction factor, if applicable), without adjusting cabinet airflow balance. Downflow nominal set point of 55 fpm was established with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI 49-2016. This cabinet model is Certified for use with optional sidewall Vacu PassTM Cord & Cable Portal. [2] Certified for use with power supplies of 230V/50 Hz and 230V/60 Hz. [3] Certified for use with power supplies of 100V/50 Hz, 100V/60 Hz, 230V/50 Hz, and 230V/60 Hz. [4] Beginning with serial number 151200002. Inflow nominal set point of 105 fpm was established with a direct airflow reading instrument. This nominal set point was confirmed using the manufacturer's recommended alternate method with thermal anemometer in a constricted access opening, using the manufacturer-supplied probe holder and sash stop (consult manufacturer for appropriate correction factor, if applicable), without adjusting cabinet airflow balance. Downflow nominal set point of 60 fpm was established with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI 49-2014. This cabinet model is Certified for use with optional sidewall Vacu-PassTM Cord & Cable Portal. [5] Inflow nominal set point of 105 fpm was established with a direct airflow reading instrument. This nominal set point was confirmed using the manufacturer's recommended alternate method with thermal anemometer in a constricted access opening, using the manufacturer-supplied probe holder and sash stop (consult manufacturer for appropriate correction factor, if applicable), without adjusting cabinet airflow balance. Downflow nominal set point of 55 fpm was established with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI 49-2018. This cabinet model is Certified for use with optional sidewall Vacu-PassTM Cord & Cable Portal. [6] Inflow nominal set point of 105 fpm was established with a direct airflow reading instrument. This nominal set point was confirmed using the manufacturer's recommended alternate method with thermal anemometer in a constricted access opening, using the manufacturer-supplied probe holder and sash stop (consult manufacturer for appropriate correction factor, if applicable), without adjusting cabinet airflow balance. Downflow nominal set point of 60 fpm was established with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI 49-2018. This cabinet model is Certified for use with optional sidewall Vacu-PassTM Cord & Cable Portal. [7] Inflow nominal set point of 105 fpm was established with a direct airflow reading instrument. This nominal set point was confirmed using the manufacturer's recommended alternate method with thermal anemometer in a constricted access opening, using the manufacturer-supplied probe holder and sash stop (consult manufacturer for appropriate correction factor, if applicable), without adjusting cabinet airflow balance. Downflow nominal set point of 65 fpm was established with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI 49-2018. This cabinet model is Certified for use with optional sidewall Vacu-PassTM Cord & Cable Portal. |
NOTE: Bench stand models 3730300, 3730310, 3401003, 3780200 and 3780100 may be used with cabinets in the 30231xxx1/30238xxx1 Logic+ series. Bench stand models 3730400, 3730410, 3401004, 3780201 and 3780101 may be used with cabinets in the 30241xxx1/30248xxx1/30242 Logic+ series and the corresponding PuriCare series. Bench stand models 3730500, 3730510, 3401005, 3780203, 3780106, 3780107, 3780312, and 3780316 may be used with cabinets in the 30251xxx1/30258xxx1 Logic+ series. Bench stand models 3730600, 3630610, 3401006, 3780202 and 3780102 may be used with cabinets in the 30261xxx1/30268xxx1/30262 Logic+ series and the corresponding PuriCare series. Bench stand models 3730400 and 34103003 may also be used with the 30348xxx1 Logic+ Series. Bench stand models 3730600 and 34103006 may also be used with the 30368xxx1 Logic+ Series. All bench stand models beginning with 378 may be used with or without the optional wheel kit (PN# 3784000).