NSF Product and Service Listings

These NSF Official Listings are current as of Saturday, July 12, 2025 at 12:15 a.m. Eastern Time. Please contact NSF to confirm the status of any Listing, report errors, or make suggestions.

Alert: NSF is concerned about fraudulent downloading and manipulation of website text. Always confirm this information by clicking on the below link for the most accurate information: http://info.nsf.org/Certified/Biosafety/Listings.asp?Company=4P720&Standard=049&


NSF/ANSI 49
Class II (Laminar Flow) Biosafety Cabinetry

Cabinet Style
A = Bench Unit With Base or Adjustable Legs Provided
B = Bench Unit Without Base or Adjustable Legs Provided
C = Console
Window Type
H = Hinged
S = Sliding
F = Fixed
Bench Height
NA = Not Applicable
NP = Not Provided

CBV = Concurrent Balance Value

Due to a change in nomenclature in NSF/ANSI 49 - 2002 “Class II (Laminar Flow) Biosafety Cabinetry,” Class II, Type A cabinets are now referred to as Class II, Type A1 and Class II, Types B3 and A/B3 cabinets as Class II, Type A2. Class II, Types B1 and B2 cabinets remain unchanged.

Biosafety cabinet models Listed under NSF/ANSI 49 are approved for use with a power supply of 115V/60Hz, unless otherwise noted. If biosafety cabinets have been approved for use with additional power supplies, the power supply will be indicated in the Official Listing.



 
Allentown, LLC 
Route 526
Allentown, NJ 08501-0698 
United States 
609-259-7951
Visit this company's website
Facility : # 1 USA

Model
Number
Cabinet
Type/Style
Inflow
Velocity
(fpm)
Downflow
Velocity
(fpm)
CBV (cfm)
at Static
Pressure
(in w.g.)
Cabinet
Width
ft.
Window
Ht/Type
in.
Bench
Ht
Max
in.
Acceptable
Options
 
39231xxx2 Safety Cab.[1] [3] A2 A 100 - 110 50 - 60 N/A 3 10S 36 Canopy Connection
I.V. Pole
U.V. Light
39232xxx2 Safety Cab.[1] [2] A2 A 100 - 110 50 - 60 N/A 3 12S 36 Canopy Connection
I.V. Pole
U.V. Light
39238xxx2 Safety Cab.[1] [3] A2 A 100 - 110 50 - 60 N/A 3 8S 36 Canopy Connection
I.V. Pole
U.V. Light
39241xxx2 Safety Cab.[1] [3] A2 A 100 - 110 50 - 60 N/A 4 10S 36 Canopy Connection
I.V. Pole
U.V. Light
39242xxx2 Safety Cab.[1] [3] A2 A 100 - 110 50 - 60 N/A 4 12S 36 Canopy Connection
I.V. Pole
U.V. Light
39248xxx2 Safety Cab.[1] [3] A2 A 100 - 110 50 - 60 N/A 4 8S 36 Canopy Connection
I.V. Pole
U.V. Light
39251xxx2 Safety Cab.[1] [3] A2 A 100 - 110 50 - 60 N/A 5 10S 36 Canopy Connection
I.V. Pole
U.V. Light
39252xxx2 Safety Cab.[1] [3] A2 A 100 - 110 50 - 60 N/A 5 12S 36 Canopy Connection
I.V. Pole
U.V. Light
39258xxx2 Safety Cab.[1] [2] A2 A 100 - 110 50 - 60 N/A 5 8S 36 Canopy Connection
I.V. Pole
U.V. Light
39261xxx2 Safety Cab.[1] [3] A2 A 100 - 110 50 - 60 N/A 6 10S 36 Canopy Connection
I.V. Pole
U.V. Light
39262xxx2 Safety Cab.[1] [4] A2 A 100 - 110 45 - 55 N/A 6 12S 36 Canopy Connection
I.V. Pole
U.V. Light
39268xxx2 Safety Cab.[1] [5] A2 A 100 - 110 50 - 60 N/A 6 8S 36 Canopy Connection
I.V. Pole
U.V. Light
39241[1] [6] A2 A 100 - 110 50 - 60 N/A 4 10S 36 Canopy Connection
I.V. Pole
U.V. Light
39242xxx1 Safety Cab.[1] [7] A2 A 100 - 110 55 - 65 N/A 4 12S 36 Canopy Connection
I.V. Pole
U.V. Light
39248[1] [6] A2 A 100 - 110 50 - 60 N/A 4 8S 36 Canopy Connection
I.V. Pole
U.V. Light
39251[1] [6] A2 A 100 - 110 50 - 60 N/A 5 10S 36 Canopy Connection
I.V. Pole
U.V. Light
39252xxx1 Safety Cab.[1] [8] A2 A 100 - 110 55 - 65 N/A 5 12S 36 Canopy Connection
I.V. Pole
U.V. Light
39258[1] [6] A2 A 100 - 110 50 - 60 N/A 5 8S 36 Canopy Connection
I.V. Pole
U.V. Light
39261[1] [6] A2 A 100 - 110 50 - 60 N/A 6 10S 36 Canopy Connection
I.V. Pole
U.V. Light
39262xxx1 Safety Cab.[1] [7] A2 A 100 - 110 55 - 65 N/A 6 12S 36 Canopy Connection
I.V. Pole
U.V. Light
39268[1] [6] A2 A 100 - 110 50 - 60 N/A 6 8S 36 Canopy Connection
I.V. Pole
U.V. Light

[1] Certified for use with power supplies of 100V/50Hz, 100V/60Hz, 230V/50Hz, and 230V/60Hz.
[2] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    measurement. This nominal set point was confirmed using the manufacturer's
    recommended alternate method with thermal anemometer in a constricted access opening,
    using the manufacturer-supplied probe holder and sash stop (consult manufacturer for
    appropriate correction factor, if applicable), without adjusting cabinet
    airflow balance. Downflow nominal set point of 55 fpm was established with I.V. pole
    and U.V. light removed. This cabinet model was Certified to NSF/ANSI 49-2022.
    This cabinet model is Certified for use with optional pre-filter, H14 filter, cable
    pass-thru, service fixtures, KD base stand, hydraulic base stand, airflow Sensor, two
    canopy models Ventus II 328202x.
[3] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument. This nominal set point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted access opening, using the
    manufacturer-supplied probe holder and sash stop (consult manufacturer for
    appropriate correction factor, if applicable), without adjusting cabinet airflow
    balance. Downflow nominal set point of 55 fpm was established with I.V. Pole and U.V.
    Light removed. This cabinet model was Certified to NSF/ANSI 49-2022. This cabinet
    model is Certified for use with optional sidewall pre-filter, H14 filter, ULPA filter,
    cable pass-thru, service fixtures, KD base stand,  hydraulic base stand, airflow Sensor,
    two canopy models Ventus II 328202x.
[4] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument. This nominal set point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted access opening, using the
    manufacturer-supplied probe holder and sash stop (consult manufacturer for appropriate
    correction factor, if applicable), without adjusting cabinet airflow balance. Downflow
    nominal set point of 50 fpm was established with I.V. Pole and U.V. Light removed.
    This cabinet model was Certified to NSF/ANSI 49-2022. This cabinet model is Certified
    for use with optional pre-filter, H14 filter, ULPA filter, canopy model Ventus II
    328202x.
[5] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    measurement. This nominal set point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted access opening,	using the
    manufacturer-supplied probe holder and sash stop (consult manufacturer for appropriate
    correction factor, if applicable), without adjusting cabinet airflow balance. Downflow
    nominal set point of 55 fpm was established with I.V. pole and U.V. light removed. This
    cabinet model was Certified to NSF/ANSI 49-2022.									
    This cabinet model is Certified for use with optional pre-filter, cable pass-thru,
    service fixtures, KD base stand, hydraulic base stand, airflow Sensor, two canopy models
    Ventus II 328202x.									
    									
    NOTE: Bench stand models 51733xx, 51713xx, 518Z3xx, 51863xx may be used with	cabinets in
    the 39238xxx2/39231xxx2/39232xxx2 SC+ series.									
     Bench stand models 51734xx, 51714xx, 518Z4xx, 51864xx may be used with cabinets in the
    39248xxx2/39241xxx2/39242xxx2  SC+ series. 	
    Bench stand models 51735xx, 51715xx, 518Z5xx, 51865xx may be used with cabinets in the
    39258xxx2/39251xxx2/39252xxx2  SC+ series.									
    Bench stand models 51736xx, 51716xx, 518Z6xx, 51866xx may be used with cabinets in the
    39268xxx2/39261xxx2/39262xxx2  SC+ series.									
    All bench stand models beginning with 518 may be used with or without the optional wheel
    kit (PN# 5190100).
[6] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    measurement. This nominal set point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted access opening, using the
    manufacturer-supplied probe holder and sash stop (consult manufacturer for appropriate
    correction factor, if applicable), without adjusting cabinet airflow balance. Downflow
    nominal set point of 55 fpm was established with I.V. pole and U.V. light removed. This
    cabinet model was Certified to NSF/ANSI 49-2012. This cabinet model is Certified for use
    with optional sidewall Vacu-PassTM Cord and Cable Portal.
[7] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    measurement. This nominal set point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted access opening, using the
    manufacturer-supplied probe holder and sash stop (consult manufacturer for appropriate
    correction factor, if applicable), without adjusting cabinet airflow balance. Downflow
    nominal set point of 55 fpm was established with I.V. pole and U.V. light removed. This
    cabinet model was Certified to NSF/ANSI 49-2018. This cabinet model is Certified for use
    with optional sidewall Vacu-PassTM Cord and Cable Portal.
[8] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument. This nominal set point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted access opening, using the
    manufacturer-supplied probe holder and sash stop (consult manufacturer for appropriate
    correction factor, if applicable), without adjusting cabinet airflow balance. Downflow
    nominal set point of 60 fpm was established with I.V. Pole and U.V. Light removed. This
    cabinet model was Certified to NSF/ANSI 49-2018. This cabinet model is Certified for use
    with optional sidewall Vacu-PassTM Cord & Cable Portal.


Number of matching Manufacturers is 1
Number of matching Products is 21