NSF Product and Service Listings

These NSF Official Listings are current as of Saturday, July 27, 2024 at 12:15 a.m. Eastern Time. Please contact NSF to confirm the status of any Listing, report errors, or make suggestions.

Alert: NSF is concerned about fraudulent downloading and manipulation of website text. Always confirm this information by clicking on the below link for the most accurate information: http://info.nsf.org/Certified/Biosafety/Listings.asp?Company=0N020&Standard=049&


NSF/ANSI 49
Class II (Laminar Flow) Biosafety Cabinetry

Cabinet Style
A = Bench Unit With Base or Adjustable Legs Provided
B = Bench Unit Without Base or Adjustable Legs Provided
C = Console
Window Type
H = Hinged
S = Sliding
F = Fixed
Bench Height
NA = Not Applicable
NP = Not Provided

CBV = Concurrent Balance Value

Due to a change in nomenclature in NSF/ANSI 49 - 2002 “Class II (Laminar Flow) Biosafety Cabinetry,” Class II, Type A cabinets are now referred to as Class II, Type A1 and Class II, Types B3 and A/B3 cabinets as Class II, Type A2. Class II, Types B1 and B2 cabinets remain unchanged.

Biosafety cabinet models Listed under NSF/ANSI 49 are approved for use with a power supply of 115V/60Hz, unless otherwise noted. If biosafety cabinets have been approved for use with additional power supplies, the power supply will be indicated in the Official Listing.



 
Esco Micro Pte Ltd 
21 Changi South Street 1
486777 
Singapore 
65 6542 0833
Visit this company's website
Facility : Kepulauan Riau, Indonesia

Model
Number
Cabinet
Type/Style
Inflow
Velocity
(fpm)
Downflow
Velocity
(fpm)
CBV (cfm)
at Static
Pressure
(in w.g.)
Cabinet
Width
ft.
Window
Ht/Type
in.
Bench
Ht
Max
in.
Acceptable
Options
 
AC2-3E8-NS G4 8"[1] [2] A2 A 100 - 110 55 - 65 N/A 3 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-3E9-NS G4 8"[1] [3] A2 A 100 - 110 55 - 65 N/A 3 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-3J7 G4 8"[1] [4] A2 A 100 - 110 55 - 65 N/A 3 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-3N7 G4 8" [1] [4] A2 A 100 - 110 55 - 65 N/A 3 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-3S8-NS G4 8"[1] [2] A2 A 100 - 110 55 - 65 N/A 3 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-3S9-NS G4 8"[1] [3] A2 A 100 - 110 55 - 65 N/A 3 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-3Y7 G4 8"[1] [4] A2 A 100 - 110 55 - 65 N/A 3 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-4E8-NS G4 10"[2] [5] A2 A 100 - 110 55 - 65 N/A 4 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-4E8-NS G4 8"[2] [5] A2 A 100 - 110 55 - 65 N/A 4 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-4E9-NS G4 10"[3] [5] A2 A 100 - 110 55 - 65 N/A 4 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-4E9-NS G4 8" [3] [5] A2 A 100 - 110 55 - 65 N/A 4 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-4J7 G4 10"[4] [5] A2 A 100 - 110 55 - 65 N/A 4 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-4J7 G4 8"[4] [5] A2 A 100 - 110 55 - 65 N/A 4 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-4N7 G4 10"[4] [5] A2 A 100 - 110 55 - 65 N/A 4 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-4N7 G4 8"[4] [5] A2 A 100 - 110 55 - 65 N/A 4 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-4S8-NS G4 10"[2] [5] A2 A 100 - 110 55 - 65 N/A 4 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-4S8-NS G4 8"[2] [5] A2 A 100 - 110 55 - 65 N/A 4 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-4S9-NS G4 10"[3] [5] A2 A 100 - 110 55 - 65 N/A 4 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-4S9-NS G4 8"[3] [5] A2 A 100 - 110 55 - 65 N/A 4 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-4Y7 G4 10"[4] [5] A2 A 100 - 110 55 - 65 N/A 4 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-4Y7 G4 8"[4] [5] A2 A 100 - 110 55 - 65 N/A 4 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-5E8-NS G4 8"[2] [6] A2 A 100 - 110 55 - 65 N/A 5 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-5E9-NS G4 8"[3] [6] A2 A 100 - 110 55 - 65 N/A 5 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-5J7 G4 8" [4] [6] A2 A 100 - 110 55 - 65 N/A 5 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-5N7 G4 8"[4] [6] A2 A 100 - 110 55 - 65 N/A 5 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-5S8-NS G4 8"[2] [6] A2 A 100 - 110 55 - 65 N/A 5 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-5S9-NS G4 8"[3] [6] A2 A 100 - 110 55 - 65 N/A 5 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-5Y7 G4 8"[4] [6] A2 A 100 - 110 55 - 65 N/A 5 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-6E8-NS G4 10"[2] [7] A2 A 100 - 110 55 - 65 N/A 6 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-6E8-NS G4 8"[2] [7] A2 A 100 - 110 55 - 65 N/A 6 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-6E9-NS G4 10"[3] [7] A2 A 100 - 110 55 - 65 N/A 6 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-6E9-NS G4 8"[3] [7] A2 A 100 - 110 55 - 65 N/A 6 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-6J7 G4 10"[4] [7] A2 A 100 - 110 55 - 65 N/A 6 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-6J7 G4 8"[4] [7] A2 A 100 - 110 55 - 65 N/A 6 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-6N7 G4 10"[4] [7] A2 A 100 - 110 55 - 65 N/A 6 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-6N7 G4 8"[4] [7] A2 A 100 - 110 55 - 65 N/A 6 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-6S8-NS G4 10" [2] [7] A2 A 100 - 110 55 - 65 N/A 6 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-6S8-NS G4 8"[2] [7] A2 A 100 - 110 55 - 65 N/A 6 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-6S9-NS G4 10"[3] [7] A2 A 100 - 110 55 - 65 N/A 6 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-6S9-NS G4 8"[3] [7] A2 A 100 - 110 55 - 65 N/A 6 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-6Y7 G4 10"[4] [7] A2 A 100 - 110 55 - 65 N/A 6 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AC2-6Y7 G4 8"[4] [7] A2 A 100 - 110 55 - 65 N/A 6 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-3E8 G4 8"[1] [2] A2 A 100 - 110 55 - 65 N/A 3 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-3E9 G4 8"[1] [3] A2 A 100 - 110 55 - 65 N/A 3 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-3S8 G4 8"[1] [2] A2 A 100 - 110 55 - 65 N/A 3 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-3S9 G4 8"[1] [3] A2 A 100 - 110 55 - 65 N/A 3 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-4E8 G4 10"[2] [5] A2 A 100 - 110 55 - 65 N/A 4 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-4E8 G4 8"[2] [5] A2 A 100 - 110 55 - 65 N/A 4 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-4E9 G4 10"[3] [5] A2 A 100 - 110 55 - 65 N/A 4 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-4E9 G4 8"[3] [5] A2 A 100 - 110 55 - 65 N/A 4 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-4S8 G4 10"[2] [5] A2 A 100 - 110 55 - 65 N/A 4 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-4S8 G4 8"[2] [5] A2 A 100 - 110 55 - 65 N/A 4 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-4S9 G4 10"[3] [5] A2 A 100 - 110 55 - 65 N/A 4 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-4S9 G4 8"[3] [5] A2 A 100 - 110 55 - 65 N/A 4 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-5E8 G4 8"[2] [6] A2 A 100 - 110 55 - 65 N/A 5 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-5E9 G4 8"[3] [6] A2 A 100 - 110 55 - 65 N/A 5 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-5S8 G4 8"[2] [6] A2 A 100 - 110 55 - 65 N/A 5 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-5S9 G4 8"[3] [6] A2 A 100 - 110 55 - 65 N/A 5 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-6E8 G4 10"[2] [7] A2 A 100 - 110 55 - 65 N/A 6 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-6E8 G4 8"[2] [7] A2 A 100 - 110 55 - 65 N/A 6 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-6E9 G4 10"[3] [7] A2 A 100 - 110 55 - 65 N/A 6 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-6E9 G4 8"[3] [7] A2 A 100 - 110 55 - 65 N/A 6 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-6S8 G4 10"[2] [7] A2 A 100 - 110 55 - 65 N/A 6 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-6S8 G4 8"[2] [7] A2 A 100 - 110 55 - 65 N/A 6 8S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-6S9 G4 10"[3] [7] A2 A 100 - 110 55 - 65 N/A 6 10S 36 I.V. Pole
U.V. Light
Canopy Connection
AR2-6S9 G4 8"[3] [7] A2 A 100 - 110 55 - 65 N/A 6 8S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-4J7 G4 10"[4] [8] A2 A 100 - 110 55 - 65 N/A 4 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-4J7 G4 12"[4] [8] A2 A 100 - 110 55 - 65 N/A 4 12S 36 I.V. Pole
U.V. Light
LA2-4J7 G4 8"[4] [8] A2 A 100 - 110 55 - 65 N/A 4 8S 36 I.V. Pole
U.V. Light
LA2-4N7 G4 10"[4] [8] A2 A 100 - 110 55 - 65 N/A 4 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-4N7 G4 12"[4] [8] A2 A 100 - 110 55 - 65 N/A 4 12S 36 I.V. Pole
U.V. Light
LA2-4N7 G4 8"[4] [8] A2 A 100 - 110 55 - 65 N/A 4 8S 36 I.V. Pole
U.V. Light
LA2-4S8 G4 10"[2] [8] A2 A 100 - 110 55 - 65 N/A 4 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-4S8 G4 12"[2] [8] A2 A 100 - 110 55 - 65 N/A 4 12S 36 I.V. Pole
U.V. Light
LA2-4S8 G4 8"[2] [8] A2 A 100 - 110 55 - 65 N/A 4 8S 36 I.V. Pole
U.V. Light
LA2-4S9 G4 10"[3] [8] A2 A 100 - 110 55 - 65 N/A 4 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-4S9 G4 12"[3] [8] A2 A 100 - 110 55 - 65 N/A 4 12S 36 I.V. Pole
U.V. Light
LA2-4S9 G4 8"[3] [8] A2 A 100 - 110 55 - 65 N/A 4 8S 36 I.V. Pole
U.V. Light
LA2-4Y7 G4 10"[4] [8] A2 A 100 - 110 55 - 65 N/A 4 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-4Y7 G4 12"[4] [8] A2 A 100 - 110 55 - 65 N/A 4 12S 36 I.V. Pole
U.V. Light
LA2-4Y7 G4 8"[4] [8] A2 A 100 - 110 55 - 65 N/A 4 8S 36 I.V. Pole
U.V. Light
LA2-5J7 G4 10"[4] [9] A2 A 100 - 110 60 - 70 N/A 5 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-5N7 G4 10"[4] [9] A2 A 100 - 110 60 - 70 N/A 5 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-5S8 G4 10"[2] [9] A2 A 100 - 110 60 - 70 N/A 5 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-5S9 G4 10"[3] [9] A2 A 100 - 110 60 - 70 N/A 5 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-5Y7 G4 10"[4] [9] A2 A 100 - 110 60 - 70 N/A 5 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-6J7 G4 10"[4] [8] A2 A 100 - 110 55 - 65 N/A 6 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-6J7 G4 12"[4] [10] A2 A 100 - 110 65 - 75 N/A 6 12S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-6N7 G4 10"[4] [8] A2 A 100 - 110 55 - 65 N/A 6 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-6N7 G4 12"[4] [10] A2 A 100 - 110 65 - 75 N/A 6 12S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-6S8 G4 10"[2] [8] A2 A 100 - 110 55 - 65 N/A 6 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-6S8 G4 12"[2] [10] A2 A 100 - 110 65 - 75 N/A 6 12S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-6S9 G4 10"[3] [8] A2 A 100 - 110 55 - 65 N/A 6 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-6S9 G4 12"[3] [10] A2 A 100 - 110 65 - 75 N/A 6 12S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-6Y7 G4 10"[4] [8] A2 A 100 - 110 55 - 65 N/A 6 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LA2-6Y7 G4 12"[4] [10] A2 A 100 - 110 65 - 75 N/A 6 12S 36 I.V. Pole
U.V. Light
Canopy Connection
LB2-4B7 G4[4] [11] B2 A 100 - 110 55 - 65 887 @ 1.9 4 8S 35 I.V. Pole
U.V. Light
Bag-In/ Bag-Out
LB2-4B8 G4[2] [11] B2 A 100 - 110 55 - 65 887 @ 1.9 4 8S 35 I.V. Pole
U.V. Light
Bag-In/ Bag-Out
LB2-4B9 G4[3] [11] B2 A 100 - 110 55 - 65 887 @ 1.9 4 8S 35 I.V. Pole
U.V. Light
Bag-In/ Bag-Out
LB2-4Y7 G4[4] [11] B2 A 100 - 110 55 - 65 887 @ 1.9 4 8S 35 I.V. Pole
U.V. Light
Bag-In/ Bag-Out
LB2-4Y8 G4[2] [11] B2 A 100 - 110 55 - 65 887 @ 1.9 4 8S 35 I.V. Pole
U.V. Light
Bag-In/ Bag-Out
LB2-4Y9 G4[3] [11] B2 A 100 - 110 55 - 65 887 @ 1.9 4 8S 35 I.V. Pole
U.V. Light
Bag-In/ Bag-Out
LB2-6B7 G4[4] [12] B2 A 100 - 110 55 - 65 1276 @ 2.7 6 8S 34 I.V. Pole
U.V. Light
Bag-In/ Bag-Out
LB2-6B8 G4[2] [12] B2 A 100 - 110 55 - 65 1276 @ 2.7 6 8S 34 I.V. Pole
U.V. Light
Bag-In/ Bag-Out
LB2-6B9 G4[3] [12] B2 A 100 - 110 55 - 65 1276 @ 2.7 6 8S 34 I.V. Pole
U.V. Light
Bag-In/ Bag-Out
LB2-6Y7 G4[4] [12] B2 A 100 - 110 55 - 65 1276 @ 2.7 6 8S 34 I.V. Pole
U.V. Light
Bag-In/ Bag-Out
LB2-6Y8 G4[2] [12] B2 A 100 - 110 55 - 65 1276 @ 2.7 6 8S 34 I.V. Pole
U.V. Light
Bag-In/ Bag-Out
LB2-6Y9 G4[3] [12] B2 A 100 - 110 55 - 65 1276 @ 2.7 6 8S 34 I.V. Pole
U.V. Light
Bag-In/ Bag-Out
LR2-4S8 G4 10"[2] [8] A2 A 100 - 110 55 - 65 N/A 4 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LR2-4S8 G4 12"[2] [8] A2 A 100 - 110 55 - 65 N/A 4 12S 36 I.V. Pole
U.V. Light
LR2-4S8 G4 8"[2] [8] A2 A 100 - 110 55 - 65 N/A 4 8S 36 I.V. Pole
U.V. Light
LR2-4S9 G4 10"[3] [8] A2 A 100 - 110 55 - 65 N/A 4 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LR2-4S9 G4 12"[3] [8] A2 A 100 - 110 55 - 65 N/A 4 12S 36 I.V. Pole
U.V. Light
LR2-4S9 G4 8"[3] [8] A2 A 100 - 110 55 - 65 N/A 4 8S 36 I.V. Pole
U.V. Light
LR2-5S8 G4 10"[2] [9] A2 A 100 - 110 55 - 65 N/A 5 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LR2-5S9 G4 10"[3] [9] A2 A 100 - 110 55 - 65 N/A 5 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LR2-6S8 G4 10"[2] [8] A2 A 100 - 110 55 - 65 N/A 6 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LR2-6S8 G4 12"[2] [10] A2 A 100 - 110 65 - 75 N/A 6 12S 36 I.V. Pole
U.V. Light
Canopy Connection
LR2-6S9 G4 10"[3] [8] A2 A 100 - 110 55 - 65 N/A 6 10S 36 I.V. Pole
U.V. Light
Canopy Connection
LR2-6S9 G4 12"[3] [10] A2 A 100 - 110 65 - 75 N/A 6 12S 36 I.V. Pole
U.V. Light
Canopy Connection
VA2-4S8 G4 12"[2] [8] A2 A 100 - 110 55 - 65 N/A 4 12S 36 I.V. Pole
U.V. Light
VA2-4S9 G4 12"[3] [8] A2 A 100 - 110 55 - 65 N/A 4 12S 36 I.V. Pole
U.V. Light
VA2-6S8 G4 12"[2] [10] A2 A 100 - 110 65 - 75 N/A 6 12S 36 I.V. Pole
U.V. Light
Canopy Connection
VA2-6S9 G4 12"[3] [10] A2 A 100 - 110 65 - 75 N/A 6 12S 36 I.V. Pole
U.V. Light
Canopy Connection

[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument. This nominal set point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted (3 inch high) access opening
    (consult manufacturer for appropriate correction factor, if applicable), without
    adjusting cabinet airflow balance.  Downflow nominal set point of 60 fpm was established
    4 inches above the bottom of the sash with the I.V. Pole and U.V. Light removed. This
    cabinet model was Certified to NSF/ANSI 49-2022.
[2] Certified for use with a power supply of 230V/50Hz and 230V/60Hz.											
[3] Certified for use with a power supply of 110V-120V/50Hz and 110V-120V/60Hz.
[4] Approved for alternate power modes of 100V/50Hz and 100V/60Hz.
[5] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument.  This nominal set point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted (3 inch high) access opening
    (consult manufacturer for appropriate correction factor, if applicable), without
    adjusting cabinet airflow balance.  Downflow nominal set point of 60 fpm was established
    with I.V. Pole and U.V. Light removed.  This cabinet model was Certified to NSF/ANSI
    49-2022.
[6] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument. This nominal set point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted (3 inch high) access opening
    (consult manufacturer for appropriate correction factor, if applicable), without
    adjusting cabinet airflow balance. Downflow nominal set point of 60 fpm was established
    4 inches above the bottom of the sash with the I.V. Pole and U.V. Light removed.  This
    cabinet model was Certified to NSF/ANSI 49-2022.
[7] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument.  This nominal set point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted (3 inch high) access opening
    (consult manufacturer for appropriate correction factor, if applicable), without
    adjusting cabinet airflow balance.  Downflow nominal set point of 60 fpm was established
    with I.V. Pole and U.V. Light removed.  This cabinet model was Certified to NSF/ANSI
    49-2022.
[8] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument. This nominal set point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted (3 inch high) access opening
    (consult manufacturer for appropriate correction factor, if applicable), without
    adjusting cabinet airflow balance. Downflow nominal set point of 60 fpm was established
    with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
    49-2014.
[9] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument. This nominal set point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted (3 inch high) access opening
    (consult manufacturer for appropriate correction factor, if applicable), without
    adjusting cabinet airflow balance. Downflow nominal set point of 65 fpm was established
    with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
    49-2014.
[10] Inflow nominal set point of 105 fpm was established with a direct airflow reading
     instrument.  This nominal set point was confirmed using the manufacturer's recommended
     alternate method with thermal anemometer in a constricted (3 inch high) access opening
     (consult manufacturer for appropriate correction factor, if applicable), without
     adjusting cabinet airflow balance. Downflow nominal set point of 70 fpm was established
     with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
     49-2014.
[11] Inflow nominal set point of 105 fpm was established with a direct airflow reading
     instrument. This nominal set point was confirmed using the manufacturer's recommended
     alternate method with thermal anemometer in a constricted (3 inch high) access opening
     (consult manufacturer for appropriate correction factor, if applicable), without
     adjusting cabinet airflow balance. Downflow nominal set point of 60 fpm was established
     with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
     49-2020.
[12] Inflow nominal set point of 105 fpm was established with a direct airflow reading
     instrument.  This nominal set point was confirmed using the manufacturer's recommended
     alternate method with thermal anemometer in a constricted (3 inch high) access opening
     (consult manufacturer for appropriate correction factor, if applicable), without
     adjusting cabinet airflow balance.  Downflow nominal set point of 60 fpm was established
     with I.V. Pole and U.V. Light removed.  This cabinet model was Certified to NSF/ANSI
     49-2022.


Number of matching Manufacturers is 1
Number of matching Products is 124
Processing time was 0 seconds