Class II (Laminar Flow) Biosafety Cabinetry
| Cabinet Style A = Bench Unit With Base or Adjustable Legs Provided B = Bench Unit Without Base or Adjustable Legs Provided C = Console |
Window Type H = Hinged S = Sliding F = Fixed |
Bench Height NA = Not Applicable NP = Not Provided |
CBV = Concurrent Balance Value
Due to a change in nomenclature in NSF/ANSI 49 - 2002 “Class II (Laminar Flow) Biosafety Cabinetry,” Class II, Type A cabinets are now referred to as Class II, Type A1 and Class II, Types B3 and A/B3 cabinets as Class II, Type A2. Class II, Types B1 and B2 cabinets remain unchanged.
Biosafety cabinet models Listed under NSF/ANSI 49 are approved for use with a power supply of 115V/60Hz, unless otherwise noted. If biosafety cabinets have been approved for use with additional power supplies, the power supply will be indicated in the Official Listing.
| Germfree Laboratories, LLC | ![]() |
| 4 Sunshine Boulevard | |
| Ormond Beach, FL 32174 | |
| United States | |
| 386-265-4360 | |
| Visit this company's website |
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| BSCA-4 | A2 | A | 100 - 110 | 54 - 64 54 - 64 40 - 50 |
N/A | 4 | 10S | 32 | I.V. Pole Canopy Connection |
[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer for appropriate correction factor, if applicable), without
adjusting cabinet airflow balance. The acceptable non-uniform downflow ranges listed
above begin with zone 1 (54-64 fpm) closest to the back wall of the cabinet, zone 2
(54-64 fpm) center of the cabinet, and end with zone 3 (40-50 fpm) closest to the front
sliding window. Downflow nominal set points were established with the I.V. Pole
removed. This cabinet model was Certified to NSF/ANSI 49-2020. This cabinet qualifies
only with the armrest installed. Certified for use with a power supply of 100V/50Hz/60Hz
and 230V/50Hz/60Hz. Additional options are I.V. pole, canopy model BSCA-4C, casters and
modified-passthrough.