NSF/ANSI 49
Class II (Laminar Flow) Biosafety Cabinetry

Cabinet Style
A = Bench Unit With Base or Adjustable Legs Provided
B = Bench Unit Without Base or Adjustable Legs Provided
C = Console
Window Type
H = Hinged
S = Sliding
F = Fixed
Bench Height
NA = Not Applicable
NP = Not Provided

CBV = Concurrent Balance Value

Due to a change in nomenclature in NSF/ANSI 49 - 2002 “Class II (Laminar Flow) Biosafety Cabinetry,” Class II, Type A cabinets are now referred to as Class II, Type A1 and Class II, Types B3 and A/B3 cabinets as Class II, Type A2. Class II, Types B1 and B2 cabinets remain unchanged.

Biosafety cabinet models Listed under NSF/ANSI 49 are approved for use with a power supply of 115V/60Hz, unless otherwise noted. If biosafety cabinets have been approved for use with additional power supplies, the power supply will be indicated in the Official Listing.



 
Germfree Laboratories, LLC 
4 Sunshine Boulevard
Ormond Beach, FL 32174 
United States 
386-265-4360
Visit this company's website
Facility : Ormond Beach, FL

Model
Number
Cabinet
Type/Style
Inflow
Velocity
(fpm)
Downflow
Velocity
(fpm)
CBV (cfm)
at Static
Pressure
(in w.g.)
Cabinet
Width
ft.
Window
Ht/Type
in.
Bench
Ht
Max
in.
Acceptable
Options
 
BSCA-4 A2 A 100 - 110 54 - 64
54 - 64
40 - 50
N/A 4 10S 32 I.V. Pole
Canopy Connection

[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument. This nominal set point was confirmed using the manufacturer's recommended 
    alternate method with thermal anemometer in a constricted (3 inch high) access opening 
    (consult manufacturer for appropriate correction factor, if applicable), without
    adjusting cabinet airflow balance. The acceptable non-uniform downflow ranges listed
    above begin with zone 1 (54-64 fpm) closest to the back wall of the cabinet, zone 2
    (54-64 fpm) center of the cabinet, and end with zone 3 (40-50 fpm) closest to the front
    sliding window.  Downflow nominal set points were established with the I.V. Pole
    removed.  This cabinet model was Certified to NSF/ANSI 49-2020.  This cabinet qualifies
    only with the armrest installed. Certified for use with a power supply of 100V/50Hz/60Hz
    and 230V/50Hz/60Hz. Additional options are I.V. pole, canopy model BSCA-4C, casters and
    modified-passthrough.