NSF Product and Service Listings

These NSF Official Listings are current as of Thursday, January 15, 2026 at 12:15 a.m. Eastern Time. Please contact NSF to confirm the status of any Listing, report errors, or make suggestions.

Alert: NSF is concerned about fraudulent downloading and manipulation of website text. Always confirm this information by clicking on the below link for the most accurate information: http://info.nsf.org/Certified/Biosafety/Listings.asp?Company=75150&Standard=049&


NSF/ANSI 49
Class II (Laminar Flow) Biosafety Cabinetry

Cabinet Style
A = Bench Unit With Base or Adjustable Legs Provided
B = Bench Unit Without Base or Adjustable Legs Provided
C = Console
Window Type
H = Hinged
S = Sliding
F = Fixed
Bench Height
NA = Not Applicable
NP = Not Provided

CBV = Concurrent Balance Value

Due to a change in nomenclature in NSF/ANSI 49 - 2002 “Class II (Laminar Flow) Biosafety Cabinetry,” Class II, Type A cabinets are now referred to as Class II, Type A1 and Class II, Types B3 and A/B3 cabinets as Class II, Type A2. Class II, Types B1 and B2 cabinets remain unchanged.

Biosafety cabinet models Listed under NSF/ANSI 49 are approved for use with a power supply of 115V/60Hz, unless otherwise noted. If biosafety cabinets have been approved for use with additional power supplies, the power supply will be indicated in the Official Listing.



 
Thermo Fisher Scientific 
Robert-Bosch-Str. 1
D-63505 Langenselbold 
Germany 
49 61 813 5277
Visit this company's website
Facility : # 2 Suzhou, China

Model
Number
Cabinet
Type/Style
Inflow
Velocity
(fpm)
Downflow
Velocity
(fpm)
CBV (cfm)
at Static
Pressure
(in w.g.)
Cabinet
Width
ft.
Window
Ht/Type
in.
Bench
Ht
Max
in.
Acceptable
Options
 
1574[1] [2] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1579[1] [2] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1584[1] [3] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1589[1] [3] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light

[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument. This nominal set-point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted (3 inch high) access opening
    (consult manufacturer for appropriate correction factor, if applicable), without
    adjusting cabinet airflow balance. Downflow nominal set-point of 63 fpm was established
    with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/
    ANSI 49-2020. Certified for use with power supplies of 230V/50Hz and 230V/60Hz.
    Removable armrests with padded and non-padded designs in addition to dished and flat
    work surface designs are approved options for this cabinet model.
[2] This model includes stainless steel interior cabinet walls with adjustable height stand.
    Model can be provided with the standard Type I power cord and internal receptacles
    (China electrical configuration with YY certification) and identified as 1574, and 1579;
    with Type GP power cord and internal receptacles (China electrical configuration
    without YY certification) and identified as 1574-GP, 1579-GP; with Type F power cord and
    internal receptacles (Korea electrical configuration) and identified as 1574-F,
    1579-F; with Type G power cord and internal receptacles (UK electrical configuration)
    and identified as 1574-G, 1579-G; with Type M power cord and internal receptacles
    (India electrical configuration) and identified as 1574-M, 1579-M; or with Type C fixed
    height stand with castors and identified as 1574-C, 1579-C, 1574-FC, 1579-FC,
    1574-GC, 1579-GC, 1574-MC, 1579-MC, 1574-C-GP, 1579-C-GP. 1546 and 1548 are the same
    models as 1574 and 1579 without stands.
[3] This model includes coated steel (SmartCoat)interior cabinet walls with adjustable
    height stand. Model can be provided with the standard Type I power cord and internal
    receptacles (China electrical configuration with YY certification) and identified as
    1584, 1589; with Type F power cord and internal receptacles (Korea electrical
    configuration)and identified as 1584-F, 1589-F; with Type G power cord and internal
    receptacles (UK electrical configuration) and identified as 1584-G, 1589-G; with Type
    M power cord and internal receptacles (India electrical configuration) and identified as
    1584-M, 1589-M; or with Type C fixed height stand with castors and identified as
    1584-C, 1589-C, 1584-FC, 1589-FC, 1584-GC, 1589-GC, 1584-MC, 1589-MC. 1556 and 1558 are
    the same models as 1584 and 1589 without stands.

Facility : Marietta, OH

Model
Number
Cabinet
Type/Style
Inflow
Velocity
(fpm)
Downflow
Velocity
(fpm)
CBV (cfm)
at Static
Pressure
(in w.g.)
Cabinet
Width
ft.
Window
Ht/Type
in.
Bench
Ht
Max
in.
Acceptable
Options
 
1305[1] [2] A2 A 100 - 110 58 - 68 N/A 4 8S 36 Armrest
I.V. Pole
U.V. Light
1306[1] [2] [3] A2 A 100 - 110 58 - 58 N/A 4 8S 36 Armrest
I.V. Pole
U.V. Light
1307[1] [4] A2 A 100 - 110 58 - 68 N/A 6 8S 36 Armrest
I.V. Pole
U.V. Light
1308[1] [3] [4] A2 A 100 - 110 58 - 68 N/A 6 8S 36 Armrest
I.V. Pole
U.V. Light
1320[5] [6] [7] A2 B 100 - 110 58 - 68 N/A 3 10S 36 Armrest
I.V. Pole
U.V. Light
1321[5] [6] [7] A2 A 100 - 110 58 - 68 N/A 3 10S 36 Armrest
I.V. Pole
U.V. Light
1322[5] [6] [7] A2 B 100 - 110 58 - 68 N/A 3 10S 36 Armrest
I.V. Pole
U.V. Light
1323[5] [6] [7] A2 A 100 - 110 58 - 68 N/A 3 10S 36 Armrest
U.V. Light
I.V. Pole
1325[1] [2] A2 A 100 - 110 58 - 68 N/A 4 8S 36 Armrest
I.V. Pole
U.V. Light
1326[1] [2] [3] A2 A 100 - 110 58 - 68 N/A 4 8S 36 Armrest
I.V. Pole
U.V. Light
1327[1] [4] A2 A 100 - 110 58 - 68 N/A 6 8S 36 Armrest
I.V. Pole
U.V. Light
1328[1] [3] [4] A2 A 100 - 110 58 - 68 N/A 6 8S 36 Armrest
I.V. Pole
U.V. Light
1330[3] [5] [6] [7] A2 B 100 - 110 58 - 68 N/A 3 10S 36 Armrest
I.V. Pole
U.V. Light
1331[3] [5] [6] [7] A2 A 100 - 110 58 - 68 N/A 3 10S 36 Armrest
I.V. Pole
U.V. Light
1332[3] [5] [6] [7] A2 B 100 - 110 58 - 68 N/A 3 10S 36 Armrest
I.V. Pole
U.V. Light
1333[3] [5] [6] [7] A2 A 100 - 110 58 - 68 N/A 3 10S 36 Armrest
I.V. Pole
U.V. Light
1335[1] [2] A2 A 100 - 110 58 - 68 N/A 4 8S 36 Armrest
I.V. Pole
U.V. Light
1336[1] [2] [3] A2 A 100 - 110 58 - 68 N/A 4 8S 36 Armrest
I.V. Pole
U.V. Light
1337[1] [4] A2 A 100 - 110 58 - 68 N/A 6 8S 36 Armrest
I.V. Pole
U.V. Light
1338[1] [3] [4] A2 A 100 - 110 58 - 68 N/A 6 8S 36 Armrest
I.V. Pole
U.V. Light
1339[5] [6] [7] A2 B 100 - 110 58 - 68 N/A 3 8S 36 Armrest
I.V. Pole
U.V. Light
1340[5] [6] [7] A2 A 100 - 110 58 - 68 N/A 3 8S 36 Armrest
I.V. Pole
U.V. Light
1341[5] [6] [7] A2 B 100 - 110 58 - 68 N/A 3 8S 36 Armrest
I.V. Pole
U.V. Light
1342[5] [6] [7] A2 A 100 - 110 58 - 68 N/A 3 8S 36 Armrest
I.V. Pole
U.V. Light
1345[2] [8] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1346[2] [8] [9] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1347[4] [8] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1348[4] [8] [9] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1350[5] [6] [10] A2 B 100 - 110 58 - 68 N/A 5 8S 36 Armrest
I.V. Pole
U.V. Light
1351[5] [6] [10] A2 A 100 - 110 58 - 68 N/A 5 8S 36 Armrest
I.V. Pole
U.V. Light
1352[5] [6] [10] A2 B 100 - 110 58 - 68 N/A 5 8S 36 Armrest
I.V. Pole
U.V. Light
1353[5] [6] [10] A2 A 100 - 110 58 - 68 N/A 5 8S 36 Armrest
I.V. Pole
U.V. Light
1354[2] [11] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1355[2] [8] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1356[2] [8] [9] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1357[4] [8] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1358[4] [8] [9] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1359[4] [8] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1366[3] [5] [6] [10] A2 B 100 - 110 58 - 68 N/A 5 8S 36 Armrest
I.V. Pole
U.V. Light
1367[3] [5] [6] [10] A2 A 100 - 110 58 - 68 N/A 5 8S 36 Armrest
I.V. Pole
U.V. Light
1368[5] [6] [10] A2 B 100 - 110 58 - 68 N/A 5 10S 36 Armrest
I.V. Pole
U.V. Light
1369[5] [6] [10] A2 A 100 - 110 58 - 68 N/A 5 10S 36 Armrest
U.V. Light
I.V. Pole
1370[5] [6] [10] A2 B 100 - 110 58 - 68 N/A 5 10S 36 Armrest
I.V. Pole
U.V. Light
1371[5] [6] [10] A2 A 100 - 110 58 - 68 N/A 5 10S 36 Armrest
I.V. Pole
U.V. Light
1372[3] [5] [6] [10] A2 B 100 - 110 58 - 68 N/A 5 8S 36 Armrest
U.V. Light
I.V. Pole
1373[3] [5] [6] [10] A2 A 100 - 110 58 - 68 N/A 5 8S 36 Armrest
I.V. Pole
U.V. Light
1375[2] [8] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1376[2] [8] [9] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1377[4] [8] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1378[4] [8] [9] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1380[3] [5] [6] [10] A2 B 100 - 110 58 - 68 N/A 5 10S 36 Armrest
I.V. Pole
U.V. Light
1381[3] [5] [6] [10] A2 A 100 - 110 58 - 68 N/A 5 10S 36 Armrest
I.V. Pole
U.V. Light
1382[3] [5] [6] [10] A2 B 100 - 110 58 - 68 N/A 5 10S 36 Armrest
I.V. Pole
U.V. Light
1383[3] [5] [6] [10] A2 A 100 - 110 58 - 68 N/A 5 10S 36 Armrest
I.V. Pole
U.V. Light
1384[2] [11] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1385[2] [8] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1386[2] [8] [9] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1387[4] [8] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1388[4] [8] [9] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1389[4] [11] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1390[3] [5] [6] [7] A2 B 100 - 110 58 - 68 N/A 3 8S 36 Armrest
I.V. Pole
U.V. Light
1391[3] [5] [6] [7] A2 A 100 - 110 58 - 68 N/A 3 8S 36 Armrest
I.V. Pole
U.V. Light
1392[3] [5] [6] [7] A2 B 100 - 110 58 - 68 N/A 3 8S 36 Armrest
I.V. Pole
U.V. Light
1393[3] [5] [6] [7] A2 A 100 - 110 58 - 68 N/A 3 8S 36 Armrest
I.V. Pole
U.V. Light
1395[2] [9] [12] A2 A 100 - 110 58 - 68 N/A 4 8S 36 Armrest
I.V. Pole
U.V. Light
1396[1] [2] [3] A2 A 100 - 110 58 - 68 N/A 4 8S 36 Armrest
I.V. Pole
U.V. Light
1397[4] [9] [12] A2 A 100 - 110 58 - 68 N/A 6 8S 36 Armrest
I.V. Pole
U.V. Light
1398[1] [3] [4] A2 A 100 - 110 58 - 68 N/A 6 8S 36 Armrest
I.V. Pole
U.V. Light
1505A2[14] [15] A2 A 100 - 110 58 - 68 N/A 4 8S 36 Armrest
I.V. Pole
U.V. Light
1506A2[13] [15] A2 A 100 - 110 58 - 68 N/A 4 8S 36 Armrest
I.V. Pole
U.V. Light
1507A2[14] [15] A2 A 100 - 110 58 - 68 N/A 6 8S 36 Armrest
I.V. Pole
U.V. Light
1508A2[13] [15] A2 A 100 - 110 58 - 68 N/A 6 8S 36 Armrest
I.V. Pole
U.V. Light
1520A2[14] [15] A2 A 100 - 110 58 - 68 N/A 3 10S 36 Armrest
I.V. Pole
U.V. Light
1521A2[14] [15] A2 A 100 - 110 58 - 68 N/A 3 10S 36 Armrest
I.V. Pole
U.V. Light
1522A2[14] [15] A2 A 100 - 110 58 - 68 N/A 3 10S 36 Armrest
I.V. Pole
U.V. Light
1523A2[14] [15] A2 A 100 - 110 58 - 68 N/A 3 10S 36 Armrest
I.V. Pole
U.V. Light
1525A2[14] [15] A2 A 100 - 110 58 - 68 N/A 4 8S 36 Armrest
I.V. Pole
U.V. Light
1526A2[13] [15] A2 A 100 - 110 58 - 68 N/A 4 8S 36 Armrest
I.V. Pole
U.V. Light
1527A2[14] [15] A2 A 100 - 110 58 - 68 N/A 6 8S 36 Armrest
I.V. Pole
U.V. Light
1528A2[13] [15] A2 A 100 - 110 58 - 68 N/A 6 8S 36 Armrest
I.V. Pole
U.V. Light
1530A2[13] [15] A2 A 100 - 110 58 - 68 N/A 3 10S 36 Armrest
I.V. Pole
U.V. Light
1531A2[13] [15] A2 A 100 - 110 58 - 68 N/A 3 10S 36 Armrest
I.V. Pole
U.V. Light
1532A2[13] [15] A2 A 100 - 110 58 - 68 N/A 3 10S 36 Armrest
I.V. Pole
U.V. Light
1533A2[13] [15] A2 A 100 - 110 58 - 68 N/A 3 10S 36 Armrest
I.V. Pole
U.V. Light
1535A2[14] [15] A2 A 100 - 110 58 - 68 N/A 4 8S 36 Armrest
I.V. Pole
U.V. Light
1536A2[13] [15] A2 A 100 - 110 58 - 68 N/A 4 8S 36 Armrest
I.V. Pole
U.V. Light
1537A2[14] [15] A2 A 100 - 110 58 - 68 N/A 6 8S 36 Armrest
I.V. Pole
U.V. Light
1538A2[13] [15] A2 A 100 - 110 58 - 68 N/A 6 8S 36 Armrest
I.V. Pole
U.V. Light
1539A2[14] [15] A2 A 100 - 110 58 - 68 N/A 3 8S 36 Armrest
I.V. Pole
U.V. Light
1540A2[14] [15] A2 A 100 - 110 58 - 68 N/A 3 8S 36 Armrest
I.V. Pole
U.V. Light
1541A2[14] [15] A2 A 100 - 110 58 - 68 N/A 3 8S 36 Armrest
I.V. Pole
U.V. Light
1542A2[14] [15] A2 A 100 - 110 58 - 68 N/A 3 8S 36 Armrest
I.V. Pole
U.V. Light
1544A2[13] [15] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1545A2[14] [15] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1547A2[14] [15] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1549A2[13] [15] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1550A2[15] [17] A2 A 100 - 110 58 - 68 N/A 5 8S 36 Armrest
I.V. Pole
U.V. Light
1551A2[15] [17] A2 A 100 - 110 58 - 68 N/A 5 8S 36 Armrest
I.V. Pole
U.V. Light
1552A2[15] [17] A2 A 100 - 110 58 - 68 N/A 5 8S 36 Armrest
I.V. Pole
U.V. Light
1553A2[15] [17] A2 A 100 - 110 58 - 68 N/A 5 8S 36 Armrest
I.V. Pole
U.V. Light
1554A2[13] [15] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1555A2[14] [15] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1557A2[14] [15] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1559A2[13] [15] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1566A2[15] [16] A2 A 100 - 110 58 - 68 N/A 5 8S 36 Armrest
I.V. Pole
U.V. Light
1567A2[15] [16] A2 A 100 - 110 58 - 68 N/A 5 8S 36 Armrest
I.V. Pole
U.V. Light
1568A2[15] [17] A2 A 100 - 110 58 - 68 N/A 5 10S 36 Armrest
I.V. Pole
U.V. Light
1569A2[15] [17] A2 A 100 - 110 58 - 68 N/A 5 10S 36 Armrest
I.V. Pole
U.V. Light
1570A2[15] [17] A2 A 100 - 110 58 - 68 N/A 5 10S 36 Armrest
I.V. Pole
U.V. Light
1571A2[15] [17] A2 A 100 - 110 58 - 68 N/A 5 10S 36 Armrest
I.V. Pole
U.V. Light
1572A2[15] [16] A2 A 100 - 110 58 - 68 N/A 5 8S 36 Armrest
I.V. Pole
U.V. Light
1573A2[15] [16] A2 A 100 - 110 58 - 68 N/A 5 8S 36 Armrest
I.V. Pole
U.V. Light
1575A2[14] [15] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1576A2[13] [15] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1577A2[14] [15] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1578A2[13] [15] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1580A2[15] [16] A2 A 100 - 110 58 - 68 N/A 5 10S 36 Armrest
I.V. Pole
U.V. Light
1581A2[15] [16] A2 A 100 - 110 58 - 68 N/A 5 10S 36 Armrest
I.V. Pole
U.V. Light
1582A2[15] [16] A2 A 100 - 110 58 - 68 N/A 5 10S 36 Armrest
I.V. Pole
U.V. Light
1583A2[15] [16] A2 A 100 - 110 58 - 68 N/A 5 10S 36 Armrest
I.V. Pole
U.V. Light
1585A2[14] [15] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1586A2[13] [15] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1587A2[14] [15] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1588A2[13] [15] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1590A2[13] [15] A2 A 100 - 110 58 - 68 N/A 3 8S 36 Armrest
I.V. Pole
U.V. Light
1591A2[13] [15] A2 A 100 - 110 58 - 68 N/A 3 8S 36 Armrest
I.V. Pole
U.V. Light
1592A2[13] [15] A2 A 100 - 110 58 - 68 N/A 3 8S 36 Armrest
I.V. Pole
U.V. Light
1593A2[13] [15] A2 A 100 - 110 58 - 68 N/A 3 8S 36 Armrest
I.V. Pole
U.V. Light
1595A2[14] [15] A2 A 100 - 110 58 - 68 N/A 4 8S 36 Armrest
I.V. Pole
U.V. Light
1596A2[13] [15] A2 A 100 - 110 58 - 68 N/A 4 8S 36 Armrest
I.V. Pole
U.V. Light
1597A2[14] [15] A2 A 100 - 110 58 - 68 N/A 6 8S 36 Armrest
I.V. Pole
U.V. Light
1598A2[13] [14] A2 A 100 - 110 58 - 68 N/A 6 8S 36 Armrest
I.V. Pole
U.V. Light

[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument. This nominal set-point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted (3 inch high) access opening
    (consult manufacturer for appropriate correction factor, if applicable), without
    adjusting cabinet airflow balance. Downflow nominal set-point of 63 fpm was established
    with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
    49-2009.
[2] Beginning April 20, 2020 a new fixed height base stand GT221201 and adjustable height
    base stand GT221189 along with 4ft cross brace GT221195 were approved for use with all
    1300 series models manufactured in Marietta OH. Casters GT221199 is an approved option
    for both stands.
[3] Certified for use with power supply of 230V/50Hz and 230V/60Hz.
[4] Beginning April 20, 2020 a new fixed height base stand GT221201 and adjustable height
    base stand GT221189 along with 6ft cross brace GT221196 were approved for use with all
    1300 series models manufactured in Marietta OH. Casters GT221199 is an approved option
    for both stands.
[5] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument.  This nominal set point was confirmed using the manufacturer's alternate
    recomended method with thermal anemometer in a constricted (3 inch high) access opening
    (consult manufacturer's operator's manual for appropriate correction factor), without
    adjusting cabinet airflow balance.  Downflow nominal set point was established with
    I.V.Pole and U.V. Light removed.  This cabinet model was Certified to NSF/ANSI 49-2009.
[6] May be used with alternate, adjustable height electric stand (maximum bench height of 35
    inches).
[7] Beginning April 20, 2020 a new fixed height base stand GT221201 and adjustable height
    base stand GT221189 along with 3ft cross brace GT221193 were approved for use with all
    1300 series models manufactured in Marietta OH. Casters GT221199 is an approved option
    for both stands.
[8] Inflow nominal set-point of 105 fpm was established with a direct airflow reading
    instrument.  This nominal set point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted (3 inch high) access opening
    (consult manufacturer for appropriate correction factor, if applicable), without
    adjusting cabinet airflow balance.  Downflow nominal set point of 63 fpm was established
    with I.V. Pole and U.V. Light removed.  This cabinet model was Certified to NSF/ANSI
    49-2009. Certified for use with power supply of 230V/50Hz and 230V/60Hz.
[9] Certified for use with power supply of 220 V / 50 Hz and 220 V / 60 Hz.
[10] Beginning April 20, 2020 a new fixed height base stand GT221201 and adjustable height
     base stand GT221189 along with 5ft cross brace GT221197 were approved for use with all
     1300 series models manufactured in Marietta OH. Casters GT221199 is an approved option
     for both stands.
[11] Inflow nominal set point of 105 fpm was established with a direct airflow reading
     instrument.  This nominal set-point was confirmed using the manufacturer's recommended
     alternate method with thermal anemometer in a constricted (3 inch high) access opening
     (consult manufacturer for appropriate correction factor, if applicable), without
     adjusting cabinet airflow balance.  Downflow nominal set-point of 63 fpm was established
     with I.V. Pole and U.V. Light removed.  This cabinet model was Certified to NSF/ANSI
     49-2009.  Certified for use with power supply of 220V/50Hz.
[12] Beginning with serial number 188122-1 and 188108-1 for models 1395 and 1397,
     respectively.  Inflow nominal set-point of 105 fpm was established with a direct airflow
     reading instrument. This nominal set point was confirmed using the manufacturer's
     recommended alternate method with thermal anemometer in a constricted (3 inch high)
     access opening (consult manufacturer for appropriate correction factor, if applicable),
     without adjusting cabinet airflow balance. Downflow nominal set point of 63 fpm was
     established with I.V. Pole and U.V. Light removed. This cabinet model was Certified to
     NSF/ANSI 49-2009.
[13] Inflow nominal set point of 105 fpm was established with a direct airflow reading
     instrument.  This nominal set point was	confirmed using the manufacturer's recommended
     alternate method with thermal anemometer in a constricted (3 inch high) access opening
     (consult manufacturer's operator's manual for appropriate correction factor) without
     adjusting cabinet airflow balance. Downflow nominal set point of 63 fpm was established
     with I.V. Pole and U.V. Light removed.  This cabinet model was Certified with both
     dished and flat work surfaces and for additional inlet power supplies of 230V/50Hz and
     230V/60Hz.  This cabinet model was Certified to NSF/ANSI 49-2022.  15xx models are
     approved for use without armrests, with	padded armrests, or with armrests without
     padding.
[14] Inflow nominal set point of 105 fpm was established with a direct airflow reading
     instrument.  This nominal set point was	confirmed using the manufacturer's recommended
     alternate method with thermal anemometer in a constricted (3 inch high) access opening
     (consult manufacturer's operator's manual for appropriate correction factor) without
     adjusting cabinet airflow balance. Downflow nominal set point of 63 fpm was established
     with I.V. Pole and U.V. Light removed.  This cabinet model was Certified with both
     dished and flat work surfaces. This cabinet model was Certified to  NSF/ANSI 49-2022. 
     15xx models are approved for use  without armrests, with  padded armrests, or with
     armrests without padding.
[15] May be used with manual adjustable height stand (maximum bench height of 35 inches),
     electrical adjustable height stand (maximum bench height of 36 inches) and caster fixed
     height stand  GT221201. Casters GT221199 is an approved option.
[16] Inflow nominal set point of 105 fpm was established with a direct airflow reading
     instrument. This nominal set point was confirmed using the manufacturer's recommended
     alternate method with thermal anemometer in a constricted (3 inch high) access opening
     (consult manufacturer's operator's manual for appropriate correction factor) without
     adjusting cabinet airflow balance. Downflow nominal set point of 63 fpm was established
     with I.V. Pole and U.V. Light removed. This cabinet model was Certified with both dished
     and flat work surfaces and for additional inlet power supplies of 230V/50Hz and
     230V/60Hz. This cabinet model was Certified to NSF/ANSI 49-2024.  15xx models are
     approved for use without armrests, with padded armrests, or with armrests without
     padding.
[17] Inflow nominal set point of 105 fpm was established with a direct airflow reading
     instrument. This nominal set point was confirmed using the manufacturer's recommended
     alternate method with thermal anemometer in a constricted (3 inch high) access opening
     (consult manufacturer's operator's manual for appropriate correction factor) without
     adjusting cabinet airflow balance. Downflow nominal set point of 63 fpm was established
     with I.V. Pole and U.V. Light removed.  This cabinet model was Certified with both
     dished and flat work surfaces. This cabinet model was Certified to NSF/ANSI 49-2024.
     15xx models are approved for use without armrests, with padded armrests, or with
     armrests without padding.

Facility : Guangzhou, China

Model
Number
Cabinet
Type/Style
Inflow
Velocity
(fpm)
Downflow
Velocity
(fpm)
CBV (cfm)
at Static
Pressure
(in w.g.)
Cabinet
Width
ft.
Window
Ht/Type
in.
Bench
Ht
Max
in.
Acceptable
Options
 
1574[1] [2] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1584[1] [3] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1579[1] [2] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1589[1] [3] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light

[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument. This nominal set-point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted (3 inch high) access opening
    (consult manufacturer for appropriate correction factor, if applicable), without
    adjusting cabinet airflow balance. Downflow nominal set-point of 63 fpm was established
    with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
    49-2020. Certified for use with power supplies of 230V/50Hz and 230V/60Hz. Removable
    armrests with padded and non-padded designs in addition to dished and flat work surface
    designs are approved options for this cabinet model.
[2] This model includes stainless steel interior cabinet walls with adjustable height stand.
    Model can be provided with the standard Type I power cord and internal receptacles
    (China electrical configuration with YY certification) and identified as 1574, and 1579;
    with Type GP power cord and internal receptacles (China electrical configuration without
    YY certification) and identified as 1574-GP, 1579-GP; with Type F power cord and
    internal receptacles (Korea electrical configuration) and identified as 1574-F, 1579-F;
    with Type G power cord and internal receptacles (UK electrical configuration) and
    identified as 1574-G, 1579-G; with Type M power cord and internal receptacles (India
    electrical configuration) and identified as 1574-M, 1579-M; or with Type C fixed height
    stand with castors and identified as 1574-C, 1579-C, 1574-FC, 1579-FC, 1574-GC, 1579-GC,
    1574-MC, 1579-MC, 1574-C-GP, 1579-C-GP. 1546 and 1548 are the same models as 1574 and
    1579 without stands.
[3] This model includes coated steel (SmartCoat)interior cabinet walls with adjustable
    height stand.  Model can be provided with the standard Type I power cord and internal
    receptacles (China electrical configuration with YY certification) and identified as
    1584, 1589; with Type F power cord and internal receptacles (Korea electrical
    configuration)and identified as 1584-F, 1589-F; with Type G power cord and internal
    receptacles (UK electrical configuration) and identified as 1584-G, 1589-G; with Type M
    power cord and internal receptacles (India electrical configuration) and identified as
    1584-M, 1589-M; or with Type C fixed height stand with castors and identified as 1584-C,
    1589-C, 1584-FC, 1589-FC, 1584-GC, 1589-GC, 1584-MC, 1589-MC. 1556 and 1558 are the same
    models as 1584 and 1589 without stands.

Facility : Guangzhou, China

Model
Number
Cabinet
Type/Style
Inflow
Velocity
(fpm)
Downflow
Velocity
(fpm)
CBV (cfm)
at Static
Pressure
(in w.g.)
Cabinet
Width
ft.
Window
Ht/Type
in.
Bench
Ht
Max
in.
Acceptable
Options
 
1574[1] [2] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1584[1] [3] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
1579[1] [2] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light
1589[1] [3] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light

[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument. This nominal set-point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted (3 inch high) access opening
    (consult manufacturer for appropriate correction factor, if applicable), without
    adjusting cabinet airflow balance. Downflow nominal set-point of 63 fpm was
    established	with I.V. Pole and U.V. Light removed. This cabinet model was Certified to
    NSF/ANSI 49-2020. Certified for use with power supplies of 230V/50Hz and 230V/60Hz.
    Removable armrests with padded and non-padded designs in addition to dished and flat
    work surface designs are approved options for this cabinet model.
[2] This model includes stainless steel interior cabinet walls with adjustable height stand.
    Model can be provided with the standard Type I power cord and internal receptacles	
    (China electrical configuration with YY certification) and identified as 1574, and 1579;
    with Type GP power cord and internal receptacles (China electrical configuration without
    YY certification) and identified as 1574-GP, 1579-GP; with Type F power cord
    and	internal receptacles (Korea electrical configuration) and identified as 1574-F,
    1579-F; with Type G power cord and internal receptacles (UK electrical configuration)
    and identified as 1574-G, 1579-G; with Type M power cord and internal receptacles (India
    electrical configuration) and identified as 1574-M, 1579-M; or with Type C fixed height
    stand with castors and identified as 1574-C, 1579-C, 1574-FC, 1579-FC, 1574-GC, 1579-GC,
    1574-MC, 1579-MC, 1574-C-GP, 1579-C-GP. 1546 and 1548 are the same models as 1574 and
    1579 without stands.
[3] This model includes coated steel (SmartCoat)interior cabinet walls with adjustable
    height stand.  Model can be provided with the standard Type I power cord and internal
    receptacles (China electrical configuration with YY certification) and identified as
    1584, 1589; with Type F power cord and internal receptacles (Korea electrical
    configuration)and identified as 1584-F, 1589-F; with Type G power cord and internal
    receptacles (UK electrical configuration) and identified as 1584-G, 1589-G; with Type
    M	power cord and internal receptacles (India electrical configuration) and identified as
    1584-M, 1589-M; or with Type C fixed height stand with castors and identified as 1584-C,
    1589-C, 1584-FC, 1589-FC, 1584-GC, 1589-GC, 1584-MC, 1589-MC. 1556 and 1558 are the same
    models as 1584 and 1589 without stands.

Facility : Langenselbold, Germany

Model
Number
Cabinet
Type/Style
Inflow
Velocity
(fpm)
Downflow
Velocity
(fpm)
CBV (cfm)
at Static
Pressure
(in w.g.)
Cabinet
Width
ft.
Window
Ht/Type
in.
Bench
Ht
Max
in.
Acceptable
Options
 
HERAsafe 2030i 1.2[1] A2 A 100 - 110 58 - 68 N/A 4 10S 36 Armrest
I.V. Pole
U.V. Light
HERAsafe 2030i 1.8[1] A2 A 100 - 110 58 - 68 N/A 6 10S 36 Armrest
I.V. Pole
U.V. Light

[1] Beginning with serial number 42294669 for the HERAsafe 2030i 1.2 and serial number
    42209172 for the HERAsafe 2030i 1.8.  Inflow nominal set point of 105 fpm was
    established with a direct airflow reading instrument.  This nominal set point was
    confirmed using the manufacturer's recommended alternate method with thermal anemometer
    in a constricted (3 inch high) access opening (consult manufacturer's operator's manual
    for appropriate correction factor) without adjusting cabinet airflow balance. Downflow
    nominal set point of 63 fpm was established with I.V. Pole and U.V. Light removed.  This
    cabinet model was Certified with both dished and flat work surfaces and for additional
    inlet power supplies of 230V/50Hz and 230V/60Hz.  This cabinet model was Certified to
    NSF/ANSI 49-2018.  HERAsafe 2030i 1.2 is approved for use with either padded armrests or
    armrests without padding.  HERAsafe 2030i 1.8 is approved for use without armrests, with
    padded armrests, or with armrests without padding.

[2] Beginning with serial number 40329642.  Inflow nominal set-point of 105 fpm was
    established with a direct airflow reading instrument.  This nominal set-point was
    confirmed using the manufacturer's recommended alternate method with thermal anemometer
    in a constricted (3 inch high) access opening (consult manufacturer's operator's manual
    for appropriate correction factor) without adjusting cabinet airflow balance. This
    cabinet model was Certified with a 3 section work surface. This cabinet model was
    Certified to NSF/ANSI 49-2009.

Number of matching Manufacturers is 1
Number of matching Products is 146