NSF Product and Service Listings
These NSF Official Listings are current as of Thursday, January 15, 2026 at 12:15 a.m. Eastern Time. Please contact NSF to confirm the status of any Listing, report errors, or make suggestions.
Alert: NSF is concerned about fraudulent downloading and manipulation of website text. Always confirm this information by clicking on the below link for the most accurate information: http://info.nsf.org/Certified/Biosafety/Listings.asp?Company=75150&Standard=049&
Class II (Laminar Flow) Biosafety Cabinetry
| Cabinet Style A = Bench Unit With Base or Adjustable Legs Provided B = Bench Unit Without Base or Adjustable Legs Provided C = Console |
Window Type H = Hinged S = Sliding F = Fixed |
Bench Height NA = Not Applicable NP = Not Provided |
CBV = Concurrent Balance Value
Due to a change in nomenclature in NSF/ANSI 49 - 2002 “Class II (Laminar Flow) Biosafety Cabinetry,” Class II, Type A cabinets are now referred to as Class II, Type A1 and Class II, Types B3 and A/B3 cabinets as Class II, Type A2. Class II, Types B1 and B2 cabinets remain unchanged.
Biosafety cabinet models Listed under NSF/ANSI 49 are approved for use with a power supply of 115V/60Hz, unless otherwise noted. If biosafety cabinets have been approved for use with additional power supplies, the power supply will be indicated in the Official Listing.
| Thermo Fisher Scientific | |
| Robert-Bosch-Str. 1 | |
| D-63505 Langenselbold | |
| Germany | |
| 49 61 813 5277 | |
| Visit this company's website |
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| 1574[1] [2] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1579[1] [2] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1584[1] [3] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1589[1] [3] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set-point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer for appropriate correction factor, if applicable), without
adjusting cabinet airflow balance. Downflow nominal set-point of 63 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/
ANSI 49-2020. Certified for use with power supplies of 230V/50Hz and 230V/60Hz.
Removable armrests with padded and non-padded designs in addition to dished and flat
work surface designs are approved options for this cabinet model.
[2] This model includes stainless steel interior cabinet walls with adjustable height stand.
Model can be provided with the standard Type I power cord and internal receptacles
(China electrical configuration with YY certification) and identified as 1574, and 1579;
with Type GP power cord and internal receptacles (China electrical configuration
without YY certification) and identified as 1574-GP, 1579-GP; with Type F power cord and
internal receptacles (Korea electrical configuration) and identified as 1574-F,
1579-F; with Type G power cord and internal receptacles (UK electrical configuration)
and identified as 1574-G, 1579-G; with Type M power cord and internal receptacles
(India electrical configuration) and identified as 1574-M, 1579-M; or with Type C fixed
height stand with castors and identified as 1574-C, 1579-C, 1574-FC, 1579-FC,
1574-GC, 1579-GC, 1574-MC, 1579-MC, 1574-C-GP, 1579-C-GP. 1546 and 1548 are the same
models as 1574 and 1579 without stands.
[3] This model includes coated steel (SmartCoat)interior cabinet walls with adjustable
height stand. Model can be provided with the standard Type I power cord and internal
receptacles (China electrical configuration with YY certification) and identified as
1584, 1589; with Type F power cord and internal receptacles (Korea electrical
configuration)and identified as 1584-F, 1589-F; with Type G power cord and internal
receptacles (UK electrical configuration) and identified as 1584-G, 1589-G; with Type
M power cord and internal receptacles (India electrical configuration) and identified as
1584-M, 1589-M; or with Type C fixed height stand with castors and identified as
1584-C, 1589-C, 1584-FC, 1589-FC, 1584-GC, 1589-GC, 1584-MC, 1589-MC. 1556 and 1558 are
the same models as 1584 and 1589 without stands.
|
|||||||||
Facility : Marietta, OH
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| 1305[1] [2] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1306[1] [2] [3] | A2 | A | 100 - 110 | 58 - 58 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1307[1] [4] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1308[1] [3] [4] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1320[5] [6] [7] | A2 | B | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1321[5] [6] [7] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1322[5] [6] [7] | A2 | B | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1323[5] [6] [7] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest U.V. Light I.V. Pole |
| 1325[1] [2] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1326[1] [2] [3] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1327[1] [4] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1328[1] [3] [4] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1330[3] [5] [6] [7] | A2 | B | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1331[3] [5] [6] [7] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1332[3] [5] [6] [7] | A2 | B | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1333[3] [5] [6] [7] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1335[1] [2] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1336[1] [2] [3] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1337[1] [4] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1338[1] [3] [4] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1339[5] [6] [7] | A2 | B | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1340[5] [6] [7] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1341[5] [6] [7] | A2 | B | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1342[5] [6] [7] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1345[2] [8] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1346[2] [8] [9] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1347[4] [8] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1348[4] [8] [9] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1350[5] [6] [10] | A2 | B | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1351[5] [6] [10] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1352[5] [6] [10] | A2 | B | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1353[5] [6] [10] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1354[2] [11] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1355[2] [8] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1356[2] [8] [9] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1357[4] [8] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1358[4] [8] [9] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1359[4] [8] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1366[3] [5] [6] [10] | A2 | B | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1367[3] [5] [6] [10] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1368[5] [6] [10] | A2 | B | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1369[5] [6] [10] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest U.V. Light I.V. Pole |
| 1370[5] [6] [10] | A2 | B | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1371[5] [6] [10] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1372[3] [5] [6] [10] | A2 | B | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest U.V. Light I.V. Pole |
| 1373[3] [5] [6] [10] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1375[2] [8] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1376[2] [8] [9] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1377[4] [8] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1378[4] [8] [9] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1380[3] [5] [6] [10] | A2 | B | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1381[3] [5] [6] [10] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1382[3] [5] [6] [10] | A2 | B | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1383[3] [5] [6] [10] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1384[2] [11] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1385[2] [8] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1386[2] [8] [9] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1387[4] [8] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1388[4] [8] [9] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1389[4] [11] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1390[3] [5] [6] [7] | A2 | B | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1391[3] [5] [6] [7] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1392[3] [5] [6] [7] | A2 | B | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1393[3] [5] [6] [7] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1395[2] [9] [12] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1396[1] [2] [3] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1397[4] [9] [12] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1398[1] [3] [4] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1505A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1506A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1507A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1508A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1520A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1521A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1522A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1523A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1525A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1526A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1527A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1528A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1530A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1531A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1532A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1533A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1535A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1536A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1537A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1538A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1539A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1540A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1541A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1542A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1544A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1545A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1547A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1549A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1550A2[15] [17] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1551A2[15] [17] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1552A2[15] [17] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1553A2[15] [17] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1554A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1555A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1557A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1559A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1566A2[15] [16] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1567A2[15] [16] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1568A2[15] [17] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1569A2[15] [17] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1570A2[15] [17] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1571A2[15] [17] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1572A2[15] [16] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1573A2[15] [16] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1575A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1576A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1577A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1578A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1580A2[15] [16] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1581A2[15] [16] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1582A2[15] [16] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1583A2[15] [16] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1585A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1586A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1587A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1588A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1590A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1591A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1592A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1593A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1595A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1596A2[13] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1597A2[14] [15] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1598A2[13] [14] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set-point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer for appropriate correction factor, if applicable), without
adjusting cabinet airflow balance. Downflow nominal set-point of 63 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
49-2009.
[2] Beginning April 20, 2020 a new fixed height base stand GT221201 and adjustable height
base stand GT221189 along with 4ft cross brace GT221195 were approved for use with all
1300 series models manufactured in Marietta OH. Casters GT221199 is an approved option
for both stands.
[3] Certified for use with power supply of 230V/50Hz and 230V/60Hz.
[4] Beginning April 20, 2020 a new fixed height base stand GT221201 and adjustable height
base stand GT221189 along with 6ft cross brace GT221196 were approved for use with all
1300 series models manufactured in Marietta OH. Casters GT221199 is an approved option
for both stands.
[5] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's alternate
recomended method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer's operator's manual for appropriate correction factor), without
adjusting cabinet airflow balance. Downflow nominal set point was established with
I.V.Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI 49-2009.
[6] May be used with alternate, adjustable height electric stand (maximum bench height of 35
inches).
[7] Beginning April 20, 2020 a new fixed height base stand GT221201 and adjustable height
base stand GT221189 along with 3ft cross brace GT221193 were approved for use with all
1300 series models manufactured in Marietta OH. Casters GT221199 is an approved option
for both stands.
[8] Inflow nominal set-point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer for appropriate correction factor, if applicable), without
adjusting cabinet airflow balance. Downflow nominal set point of 63 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
49-2009. Certified for use with power supply of 230V/50Hz and 230V/60Hz.
[9] Certified for use with power supply of 220 V / 50 Hz and 220 V / 60 Hz.
[10] Beginning April 20, 2020 a new fixed height base stand GT221201 and adjustable height
base stand GT221189 along with 5ft cross brace GT221197 were approved for use with all
1300 series models manufactured in Marietta OH. Casters GT221199 is an approved option
for both stands.
[11] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set-point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer for appropriate correction factor, if applicable), without
adjusting cabinet airflow balance. Downflow nominal set-point of 63 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
49-2009. Certified for use with power supply of 220V/50Hz.
[12] Beginning with serial number 188122-1 and 188108-1 for models 1395 and 1397,
respectively. Inflow nominal set-point of 105 fpm was established with a direct airflow
reading instrument. This nominal set point was confirmed using the manufacturer's
recommended alternate method with thermal anemometer in a constricted (3 inch high)
access opening (consult manufacturer for appropriate correction factor, if applicable),
without adjusting cabinet airflow balance. Downflow nominal set point of 63 fpm was
established with I.V. Pole and U.V. Light removed. This cabinet model was Certified to
NSF/ANSI 49-2009.
[13] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer's operator's manual for appropriate correction factor) without
adjusting cabinet airflow balance. Downflow nominal set point of 63 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model was Certified with both
dished and flat work surfaces and for additional inlet power supplies of 230V/50Hz and
230V/60Hz. This cabinet model was Certified to NSF/ANSI 49-2022. 15xx models are
approved for use without armrests, with padded armrests, or with armrests without
padding.
[14] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer's operator's manual for appropriate correction factor) without
adjusting cabinet airflow balance. Downflow nominal set point of 63 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model was Certified with both
dished and flat work surfaces. This cabinet model was Certified to NSF/ANSI 49-2022.
15xx models are approved for use without armrests, with padded armrests, or with
armrests without padding.
[15] May be used with manual adjustable height stand (maximum bench height of 35 inches),
electrical adjustable height stand (maximum bench height of 36 inches) and caster fixed
height stand GT221201. Casters GT221199 is an approved option.
[16] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer's operator's manual for appropriate correction factor) without
adjusting cabinet airflow balance. Downflow nominal set point of 63 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model was Certified with both dished
and flat work surfaces and for additional inlet power supplies of 230V/50Hz and
230V/60Hz. This cabinet model was Certified to NSF/ANSI 49-2024. 15xx models are
approved for use without armrests, with padded armrests, or with armrests without
padding.
[17] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer's operator's manual for appropriate correction factor) without
adjusting cabinet airflow balance. Downflow nominal set point of 63 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model was Certified with both
dished and flat work surfaces. This cabinet model was Certified to NSF/ANSI 49-2024.
15xx models are approved for use without armrests, with padded armrests, or with
armrests without padding.
|
|||||||||
Facility : Guangzhou, China
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| 1574[1] [2] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1584[1] [3] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1579[1] [2] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1589[1] [3] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set-point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer for appropriate correction factor, if applicable), without
adjusting cabinet airflow balance. Downflow nominal set-point of 63 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
49-2020. Certified for use with power supplies of 230V/50Hz and 230V/60Hz. Removable
armrests with padded and non-padded designs in addition to dished and flat work surface
designs are approved options for this cabinet model.
[2] This model includes stainless steel interior cabinet walls with adjustable height stand.
Model can be provided with the standard Type I power cord and internal receptacles
(China electrical configuration with YY certification) and identified as 1574, and 1579;
with Type GP power cord and internal receptacles (China electrical configuration without
YY certification) and identified as 1574-GP, 1579-GP; with Type F power cord and
internal receptacles (Korea electrical configuration) and identified as 1574-F, 1579-F;
with Type G power cord and internal receptacles (UK electrical configuration) and
identified as 1574-G, 1579-G; with Type M power cord and internal receptacles (India
electrical configuration) and identified as 1574-M, 1579-M; or with Type C fixed height
stand with castors and identified as 1574-C, 1579-C, 1574-FC, 1579-FC, 1574-GC, 1579-GC,
1574-MC, 1579-MC, 1574-C-GP, 1579-C-GP. 1546 and 1548 are the same models as 1574 and
1579 without stands.
[3] This model includes coated steel (SmartCoat)interior cabinet walls with adjustable
height stand. Model can be provided with the standard Type I power cord and internal
receptacles (China electrical configuration with YY certification) and identified as
1584, 1589; with Type F power cord and internal receptacles (Korea electrical
configuration)and identified as 1584-F, 1589-F; with Type G power cord and internal
receptacles (UK electrical configuration) and identified as 1584-G, 1589-G; with Type M
power cord and internal receptacles (India electrical configuration) and identified as
1584-M, 1589-M; or with Type C fixed height stand with castors and identified as 1584-C,
1589-C, 1584-FC, 1589-FC, 1584-GC, 1589-GC, 1584-MC, 1589-MC. 1556 and 1558 are the same
models as 1584 and 1589 without stands.
|
|||||||||
Facility : Guangzhou, China
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| 1574[1] [2] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1584[1] [3] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1579[1] [2] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1589[1] [3] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set-point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer for appropriate correction factor, if applicable), without
adjusting cabinet airflow balance. Downflow nominal set-point of 63 fpm was
established with I.V. Pole and U.V. Light removed. This cabinet model was Certified to
NSF/ANSI 49-2020. Certified for use with power supplies of 230V/50Hz and 230V/60Hz.
Removable armrests with padded and non-padded designs in addition to dished and flat
work surface designs are approved options for this cabinet model.
[2] This model includes stainless steel interior cabinet walls with adjustable height stand.
Model can be provided with the standard Type I power cord and internal receptacles
(China electrical configuration with YY certification) and identified as 1574, and 1579;
with Type GP power cord and internal receptacles (China electrical configuration without
YY certification) and identified as 1574-GP, 1579-GP; with Type F power cord
and internal receptacles (Korea electrical configuration) and identified as 1574-F,
1579-F; with Type G power cord and internal receptacles (UK electrical configuration)
and identified as 1574-G, 1579-G; with Type M power cord and internal receptacles (India
electrical configuration) and identified as 1574-M, 1579-M; or with Type C fixed height
stand with castors and identified as 1574-C, 1579-C, 1574-FC, 1579-FC, 1574-GC, 1579-GC,
1574-MC, 1579-MC, 1574-C-GP, 1579-C-GP. 1546 and 1548 are the same models as 1574 and
1579 without stands.
[3] This model includes coated steel (SmartCoat)interior cabinet walls with adjustable
height stand. Model can be provided with the standard Type I power cord and internal
receptacles (China electrical configuration with YY certification) and identified as
1584, 1589; with Type F power cord and internal receptacles (Korea electrical
configuration)and identified as 1584-F, 1589-F; with Type G power cord and internal
receptacles (UK electrical configuration) and identified as 1584-G, 1589-G; with Type
M power cord and internal receptacles (India electrical configuration) and identified as
1584-M, 1589-M; or with Type C fixed height stand with castors and identified as 1584-C,
1589-C, 1584-FC, 1589-FC, 1584-GC, 1589-GC, 1584-MC, 1589-MC. 1556 and 1558 are the same
models as 1584 and 1589 without stands.
|
|||||||||
Facility : Langenselbold, Germany
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| HERAsafe 2030i 1.2[1] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| HERAsafe 2030i 1.8[1] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
[1] Beginning with serial number 42294669 for the HERAsafe 2030i 1.2 and serial number
42209172 for the HERAsafe 2030i 1.8. Inflow nominal set point of 105 fpm was
established with a direct airflow reading instrument. This nominal set point was
confirmed using the manufacturer's recommended alternate method with thermal anemometer
in a constricted (3 inch high) access opening (consult manufacturer's operator's manual
for appropriate correction factor) without adjusting cabinet airflow balance. Downflow
nominal set point of 63 fpm was established with I.V. Pole and U.V. Light removed. This
cabinet model was Certified with both dished and flat work surfaces and for additional
inlet power supplies of 230V/50Hz and 230V/60Hz. This cabinet model was Certified to
NSF/ANSI 49-2018. HERAsafe 2030i 1.2 is approved for use with either padded armrests or
armrests without padding. HERAsafe 2030i 1.8 is approved for use without armrests, with
padded armrests, or with armrests without padding.
|
|||||||||
[2] Beginning with serial number 40329642. Inflow nominal set-point of 105 fpm was
established with a direct airflow reading instrument. This nominal set-point was
confirmed using the manufacturer's recommended alternate method with thermal anemometer
in a constricted (3 inch high) access opening (consult manufacturer's operator's manual
for appropriate correction factor) without adjusting cabinet airflow balance. This
cabinet model was Certified with a 3 section work surface. This cabinet model was
Certified to NSF/ANSI 49-2009.
Number of matching Manufacturers is 1
Number of matching Products is 146
Processing time was 0 seconds