NSF Product and Service Listings

These NSF Official Listings are current as of Thursday, January 15, 2026 at 12:15 a.m. Eastern Time. Please contact NSF to confirm the status of any Listing, report errors, or make suggestions.

Alert: NSF is concerned about fraudulent downloading and manipulation of website text. Always confirm this information by clicking on the below link for the most accurate information: http://info.nsf.org/Certified/Biosafety/Listings.asp?Company=C0000779&Standard=049&


NSF/ANSI 49
Class II (Laminar Flow) Biosafety Cabinetry

Cabinet Style
A = Bench Unit With Base or Adjustable Legs Provided
B = Bench Unit Without Base or Adjustable Legs Provided
C = Console
Window Type
H = Hinged
S = Sliding
F = Fixed
Bench Height
NA = Not Applicable
NP = Not Provided

CBV = Concurrent Balance Value

Due to a change in nomenclature in NSF/ANSI 49 - 2002 “Class II (Laminar Flow) Biosafety Cabinetry,” Class II, Type A cabinets are now referred to as Class II, Type A1 and Class II, Types B3 and A/B3 cabinets as Class II, Type A2. Class II, Types B1 and B2 cabinets remain unchanged.

Biosafety cabinet models Listed under NSF/ANSI 49 are approved for use with a power supply of 115V/60Hz, unless otherwise noted. If biosafety cabinets have been approved for use with additional power supplies, the power supply will be indicated in the Official Listing.



 
Thermo Fisher Scientific 
401 Millcreek Road
Marietta, OH 45750 
United States 
740-373-4763
Facility : # 1 USA

Model
Number
Cabinet
Type/Style
Inflow
Velocity
(fpm)
Downflow
Velocity
(fpm)
CBV (cfm)
at Static
Pressure
(in w.g.)
Cabinet
Width
ft.
Window
Ht/Type
in.
Bench
Ht
Max
in.
Acceptable
Options
 
1501B2[1] [2] B2 A 100 - 110 50 - 60 818 @ 1.8 4 8S 36 I.V. Pole
U.V. Light
1502B2[1] [2] B2 A 100 - 110 50 - 60 818 @ 1.8 4 8S 36 I.V. Pole
U.V. Light
1503B2[1] [2] B2 A 100 - 110 50 - 60 818 @ 1.8 4 8S 36 I.V. Pole
U.V. Light
1504B2[1] [2] B2 A 100 - 110 50 - 60 818 @ 1.8 4 8S 36 I.V. Pole
U.V. Light
1513B2[1] [2] B2 A 100 - 110 50 - 60 818 @ 1.8 4 8S 36 I.V. Pole
U.V. Light
1514B2[1] [2] B2 A 100 - 110 50 - 60 818 @ 1.8 4 8S 36 I.V. Pole
U.V. Light
1513B2F[1] [2] B2 A 100 - 110 50 - 60 818 @ 1.8 4 8S 36 I.V. Pole
U.V. Light
1513B2G[1] [2] B2 A 100 - 110 50 - 60 818 @ 1.8 4 8S 36 I.V. Pole
U.V. Light
1513B2M [1] [2] B2 A 100 - 110 50 - 60 818 @ 1.8 4 8S 36 I.V. Pole
U.V. Light
1514B2F[1] [2] B2 A 100 - 110 50 - 60 818 @ 1.8 4 8S 36 I.V. Pole
U.V. Light
1514B2G[1] [2] B2 A 100 - 110 50 - 60 818 @ 1.8 4 8S 36 I.V. Pole
U.V. Light
1514B2M[1] [2] B2 A 100 - 110 50 - 60 818 @ 1.8 4 8S 36 I.V. Pole
U.V. Light
1512B2[1] [2] B2 A 100 - 110 50 - 60 1382 @ 2.5 6 8S 36 I.V. Pole
U.V. Light
1519B2[1] [2] B2 A 100 - 110 50 - 60 1382 @ 2.5 6 8S 36 I.V. Pole
U.V. Light
1534B2[1] [2] B2 A 100 - 110 50 - 60 1382 @ 2.5 6 8S 36 I.V. Pole
U.V. Light
1543B2[1] [2] B2 A 100 - 110 50 - 60 1382 @ 2.5 6 8S 36 I.V. Pole
U.V. Light
1524B2[1] [2] B2 A 100 - 110 50 - 60 1382 @ 2.5 6 8S 36 I.V. Pole
U.V. Light
1529B2[1] [2] B2 A 100 - 110 50 - 60 1382 @ 2.5 6 8S 36 I.V. Pole
U.V. Light
1524B2F[1] [2] B2 A 100 - 110 50 - 60 1382 @ 2.5 6 8S 36 I.V. Pole
U.V. Light
1524B2G[1] [2] B2 A 100 - 110 50 - 60 1382 @ 2.5 6 8S 36 I.V. Pole
U.V. Light
1524B2M[1] [2] B2 A 100 - 110 50 - 60 1382 @ 2.5 6 8S 36 I.V. Pole
U.V. Light
1529B2F[1] [2] B2 A 100 - 110 50 - 60 1382 @ 2.5 6 8S 36 I.V. Pole
U.V. Light
1529B2G[1] [2] B2 A 100 - 110 50 - 60 1382 @ 2.5 6 8S 36 I.V. Pole
U.V. Light
1529B2M[1] [2] B2 A 100 - 110 50 - 60 1382 @ 2.5 6 8S 36 I.V. Pole
U.V. Light
151x[2] [3] [4] B2 A 100 - 110 50 - 60 852 @ 1.8 4 8S 36 Bag-In/ Bag-Out
I.V. Pole
U.V. Light
156x[2] [3] [5] B2 A 100 - 110 50 - 60 1265 @ 2.2 6 8S 36 Bag-In/ Bag-Out
I.V. Pole
U.V. Light

[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument.  This nominal set point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted access opening, using the
    manufacturer-supplied probe holder and sash stop (consult manufacturer for appropriate
    correction factor, if applicable), without adjusting cabinet airflow balance. Downflow
    nominal set point of 55 fpm was established with I.V. Pole and U.V. Light removed.  This
    cabinet model was Certified to NSF/ANSI 49-2022.  This cabinet model is Certified for
    use with optional pre-filter, cable pass-Thru, service fixtures, welded stand 3730402
    (4ft) and 3730602 (6ft), and airflow Sensor.
[2] Certified for use with power supplies of 100V/50 Hz, 100V/60 Hz, 230V/50 Hz, and 230V/60
    Hz.
[3] Beginning with serial number 190500000. Inflow nominal set point of 105 fpm was
    established with a direct airflow reading instrument.  This nominal set point was
    confirmed using the manufacturer's recommended alternate method with thermal anemometer
    in a constricted access opening, using the manufacturer-supplied probe holder and sash
    stop (consult manufacturer for appropriate correction factor, if applicable), without
    adjusting cabinet airflow balance. Downflow nominal set point of 55 fpm was established
    with I.V. Pole and U.V. Light removed.  This cabinet model was Certified to NSF/ANSI
    49-2016.  This cabinet model is Certified for use with optional cord & cable portal.
[4] Model 1511 is equivalent to Model 1510, but comes equipped with U.V. lighting (Listed as
    an acceptable option).  Model 1515 is equivalent to model 1510, but is rated for use
    with power supplies of 230V and 50/60 Hz and is equipped with a Chinese-compatible plug.
     Model 1516 is equivalent, but is rated for use with power supplies of 230V and 50/60 Hz
    and is equipped with U.V. lighting and a Chinese-compatible plug. Model 1517 is
    equivalent, but is rated for use with power supplies of 230V and 50/60 Hz and is
    equipped with a North American 230 Volt-compatible plug. Model 1517F is equivalent, but
    is rated for use with power supplies of 230V and 50/60 Hz and is equipped with a Type F
    electrical plug (Schuko).  Model 1517G is equivalent, but is rated for use with power
    supplies of 230V and 50/60 Hz and is equipped with a Type G electrical plug
    (UK-compatible). Model 1517M is equivalent, but is rated for use with power supplies of
    230V and 50/60 Hz and is equipped with a Type M electrical plug (India-compatible). 
    Model 1518 is equivalent, but is rated for use with power supplies of 230V and 50/60 Hz
    and is equipped with U.V. lighting and a North American 230 Volt-compatible plug.Model
    1518F is equivalent, but is rated for use with power supplies of 230V and 50/60 Hz and
    is equipped with a Type F electrical plug (Schuko).  Model 1518G is equivalent, but is
    rated for use with power supplies of 230V and 50/60 Hz and is equipped with a Type G
    electrical plug (UK-compatible).  Model 1518M is equivalent, but is rated for use with
    power supplies of 230V and 50/60 Hz and is equipped with a Type M electrical plug
    (India-compatible).
[5] Model 1561 is equivalent to Model 1560, but comes equipped with U.V. lighting (Listed as
    an acceptable option).  Model 1562 is equivalent to model 1560, but is rated for use
    with power supplies of 230V and 50/60 Hz and is equipped with a Chinese-compatible plug.
     Model 1563 is equivalent, but is rated for use with power supplies of 230V and 50/60 Hz
    and is equipped with U.V. lighting and a Chinese-compatible plug. Model 1564 is
    equivalent, but is rated for use with power supplies of 230V and 50/60 Hz and is
    equipped with a North American 230 Volt-compatible plug.  Model 1564F is equivalent, but
    is rated for use with power supplies of 230V and 50/60 Hz and is equipped with a Type F
    electrical plug (Schuko).  Model 1564G is equivalent, but is rated for use with power
    supplies of 230V and 50/60 Hz and is equipped with a Type G electrical plug
    (UK-compatible).  Model 1564M is equivalent, but is rated for use with power supplies of
    230V and 50/60 Hz and is equipped with a Type M electrical plug (India-compatible).
    Model 1565 is equivalent, but is rated for use with power supplies of 230 V and 50/60 Hz
    and is equipped with U.V. lighting and a North American 230 Volt-compatible plug. Model
    1565F is equivalent, but is rated for use with power supplies of 230V and 50/60 Hz and
    is equipped with a Type F electrical plug (Schuko). Model 1565G is equivalent, but is
    rated for use with power supplies of 230V and 50/60 Hz and is equipped with a Type G
    electrical plug (UK-compatible).  Model 1565M is equivalent, but is rated for use with
    power supplies of 230V and 50/60 Hz and is equipped with a Type M electrical plug
    (India-compatible).


Number of matching Manufacturers is 1
Number of matching Products is 26
Processing time was 0 seconds