Class II (Laminar Flow) Biosafety Cabinetry
| Cabinet Style A = Bench Unit With Base or Adjustable Legs Provided B = Bench Unit Without Base or Adjustable Legs Provided C = Console |
Window Type H = Hinged S = Sliding F = Fixed |
Bench Height NA = Not Applicable NP = Not Provided |
CBV = Concurrent Balance Value
Due to a change in nomenclature in NSF/ANSI 49 - 2002 “Class II (Laminar Flow) Biosafety Cabinetry,” Class II, Type A cabinets are now referred to as Class II, Type A1 and Class II, Types B3 and A/B3 cabinets as Class II, Type A2. Class II, Types B1 and B2 cabinets remain unchanged.
Biosafety cabinet models Listed under NSF/ANSI 49 are approved for use with a power supply of 115V/60Hz, unless otherwise noted. If biosafety cabinets have been approved for use with additional power supplies, the power supply will be indicated in the Official Listing.
| Design Filtration Microzone Inc. | |
| 86 Harry Douglas Drive | |
| Ottawa, ON K2S 2C7 | |
| Canada | |
| 613-831-8318 | |
| Visit this company's website |
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| BK-3-4[1] | A2 | A | 100 - 110 | 60 - 70 | N/A | 4 | 10S | 27 | Canopy Connection I.V. Pole U.V. Light |
| BK-3-6[2] | A2 | A | 100 - 110 | 60 - 70 | N/A | 6 | 10S | 27 | I.V. Pole U.V. Light Canopy Connection |
| BK3-4[3] | A2 | A | 100 - 110 | 60 - 70 | N/A | 4 | 8S | 27 | I.V. Pole U.V. Light |
[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. A corresponding inflow nominal set point of 105 fpm (corrected to local
air density) was confirmed using manufacturer's recommended alternate method with
thermal anemometer in a constricted (3 inch high) access opening and applying
appropriate correction factor without adjusting cabinet airflow balance. The downflow
nominal set point of 65 fpm was established 4 inches above the bottom of the sash with
the I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
49-2014. Certified for use with power supplies of 115V/60Hz, 230V/50Hz and 230V/60Hz.
Canopy connection model number is 4FT-CC.
[2] Inflow nominal set-point of 105 fpm was established with a direct airflow reading
instrument. A corresponding inflow nominal set-point of 105 fpm (corrected to local air
density) was confirmed using manufacturer's recommended alternate method with thermal
anemometer in a constricted (3 inch high) access opening and applying appropriate
correction factor without adjusting cabinet airflow balance. The downflow nominal
set-point of 65 fpm was
established 4 inches above the bottom of the sash with the I.V. Pole and U.V. Light
removed. This cabinet model was Certified to NSF/ANSI 49-2014. Certified for use with
power supplies of 115V/60Hz, 230V/50Hz and 230V/60Hz. Canopy connection model number is
6FT-CC.
[3] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set-point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer for appropriate correction factor, if applicable), without
adjusting cabinet airflow balance. Downflow nominal set-point of 65 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
49-2012. Approved for alternate power modes of 115V/60Hz, 220V/50Hz and 220V/60Hz.
Canopy connection model number is 4FE-CC.
|
|||||||||