NSF/ANSI 49
Class II (Laminar Flow) Biosafety Cabinetry

Cabinet Style
A = Bench Unit With Base or Adjustable Legs Provided
B = Bench Unit Without Base or Adjustable Legs Provided
C = Console
Window Type
H = Hinged
S = Sliding
F = Fixed
Bench Height
NA = Not Applicable
NP = Not Provided

CBV = Concurrent Balance Value

Due to a change in nomenclature in NSF/ANSI 49 - 2002 “Class II (Laminar Flow) Biosafety Cabinetry,” Class II, Type A cabinets are now referred to as Class II, Type A1 and Class II, Types B3 and A/B3 cabinets as Class II, Type A2. Class II, Types B1 and B2 cabinets remain unchanged.

Biosafety cabinet models Listed under NSF/ANSI 49 are approved for use with a power supply of 115V/60Hz, unless otherwise noted. If biosafety cabinets have been approved for use with additional power supplies, the power supply will be indicated in the Official Listing.



 
Mystaire Inc. 
1200 Telecom Drive
Creedmoor, NC 27522 
United States 
919-229-8511
Visit this company's website
Facility : Shandong Province, China

Model
Number
Cabinet
Type/Style
Inflow
Velocity
(fpm)
Downflow
Velocity
(fpm)
CBV (cfm)
at Static
Pressure
(in w.g.)
Cabinet
Width
ft.
Window
Ht/Type
in.
Bench
Ht
Max
in.
Acceptable
Options
 
MY-24A2[1] [2] A2 A 100 - 110 60 - 70 N/A 2 10S 28 U.V. Light
Canopy Connection
MY-36A2[2] [3] A2 A 100 - 110 60 - 70 N/A 3 10S 27 U.V. Light
Canopy Connection
I.V. Pole
MY-48A2[2] [4] A2 A 100 - 110 60 - 70 N/A 4 10S 27 U.V. Light
Canopy Connection
I.V. Pole
MY-48B2[2] [5] B2 A 100 - 110 60 - 70 949 @ 2.1 4 8S 27 U.V. Light
I.V. Pole
MY-72A2[2] [6] A2 A 100 - 110 60 - 70 N/A 6 10S 27 U.V. Light
Canopy Connection
I.V. Pole

[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument.  This nominal set point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted (3 inch high) access opening
    (consult manufacturer's operator's manual for appropriate correction factor) without
    adjusting cabinet airflow balance. Downflow nominal set point of 65 fpm was established
    with I.V. Pole and U.V. Light removed.  This cabinet model was Certified to NSF/ANSI
    49-2020. Canopy connection model number is 2FT-CC.
[2] Certified for use with power supplies of 230V/50Hz and 230V/60Hz.
[3] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument.  This nominal set point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted (3 inch high) access opening
    (consult manufacturer's operator's manual for appropriate correction factor) without
    adjusting cabinet airflow balance. Downflow nominal set point of 65 fpm was established
    with I.V. Pole and U.V. Light removed.  This cabinet model was Certified to NSF/ANSI
    49-2016. Canopy model 3FT-CC was evaluated as a part of the NSF Certification testing
    procedures.
[4] Inflow nominal set-point of 105 fpm was established with a direct airflow reading
    instrument. A corresponding inflow nominal set-point of 105 fpm (corrected to local air
    density) was confirmed using manufacturer's recommended alternate method with thermal
    anemometer in a constricted (3 inch high) access opening and applying appropriate
    correction factor without adjusting cabinet airflow balance. The downflow nominal
    set-point of 65 fpm was established 4 inches above the bottom of the sash with the I.V.
    Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI 49-2014.
    Canopy model 4FT-CC was evaluated as a part of the NSF Certification testing procedures.
[5] Inflow nominal set point of 105 fpm was established with a direct airflow reading
    instrument.  This nominal set point was confirmed using the manufacturer's recommended
    alternate method with thermal anemometer in a constricted (3 inch high) access opening
    (consult manufacturer's operator's manual for appropriate correction factor) without
    adjusting cabinet airflow balance. Downflow nominal set point of 65 fpm was established
    with I.V. Pole and U.V. Light removed.  This cabinet model was Certified to NSF/ANSI
    49-2016.
[6] Inflow nominal set-point of 105 fpm was established with a direct airflow reading
    instrument.  A corresponding inflow nominal set-point of 105 fpm (corrected to local air
    density) was confirmed using manufacturer's recommended alternate method with thermal
    anemometer in a constricted (3 inch high) access opening and applying appropriate
    correction factor without adjusting cabinet airflow balance.  The downflow nominal
    set-point of 65 fpm was established 4 inches above the bottom of the sash with the I.V.
    Pole and U.V. Light removed.  This cabinet model was Certified to NSF/ANSI 49-2014.
    Canopy model 6FT-CC was evaluated as a part of the NSF Certification testing procedures.


Number of matching Manufacturers is 1
Number of matching Products is 5