Class II (Laminar Flow) Biosafety Cabinetry
| Cabinet Style A = Bench Unit With Base or Adjustable Legs Provided B = Bench Unit Without Base or Adjustable Legs Provided C = Console |
Window Type H = Hinged S = Sliding F = Fixed |
Bench Height NA = Not Applicable NP = Not Provided |
CBV = Concurrent Balance Value
Due to a change in nomenclature in NSF/ANSI 49 - 2002 “Class II (Laminar Flow) Biosafety Cabinetry,” Class II, Type A cabinets are now referred to as Class II, Type A1 and Class II, Types B3 and A/B3 cabinets as Class II, Type A2. Class II, Types B1 and B2 cabinets remain unchanged.
Biosafety cabinet models Listed under NSF/ANSI 49 are approved for use with a power supply of 115V/60Hz, unless otherwise noted. If biosafety cabinets have been approved for use with additional power supplies, the power supply will be indicated in the Official Listing.
| iGene Labserve Private Limited | |
| 16/2, Ashok Nagar | |
| New Delhi | |
| 110018 | |
| India | |
| 91 11 79613060 |
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| IG-CIIA222G[1] [5] | A2 | A | 100 - 110 | 60 - 70 | N/A | 2 | 10S | 28 | U.V. Light Canopy Connection |
| IG-CIIA222N[1] [5] | A2 | A | 100 - 110 | 60 - 70 | N/A | 2 | 10S | 28 | U.V. Light Canopy Connection |
| IG-CIIA242G[2] [5] | A2 | A | 100 - 110 | 60 - 70 | N/A | 4 | 10S | 27 | U.V. Light Canopy Connection I.V. Pole |
| IG-CIIA242N[2] [5] | A2 | A | 100 - 110 | 60 - 70 | N/A | 4 | 10S | 27 | U.V. Light Canopy Connection I.V. Pole |
| IG-CIIB242G[3] [5] | B2 | A | 100 - 110 | 60 - 70 | 949 @ 2.1 | 4 | 8S | 27 | U.V. Light I.V. Pole |
| IG-CIIB242N[3] [5] | B2 | A | 100 - 110 | 60 - 70 | 949 @ 2.1 | 4 | 8S | 27 | U.V. Light I.V. Pole |
| IG-CIIA262G[4] [5] | A2 | A | 100 - 110 | 60 - 70 | N/A | 6 | 10S | 27 | U.V. Light Canopy Connection I.V. Pole |
| IG-CIIA262N[4] [5] | A2 | A | 100 - 110 | 60 - 70 | N/A | 6 | 10S | 27 | U.V. Light Canopy Connection I.V. Pole |
[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer's operator's manual for appropriate correction factor) without
adjusting cabinet airflow balance. Downflow nominal set point of 65 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
49-2020. Certified canopy connection model is 2FT-CC.
[2] Inflow nominal set-point of 105 fpm was established with a direct airflow reading
instrument. A corresponding inflow nominal set-point of 105 fpm (corrected to local air
density) was confirmed using manufacturer's recommended alternate method with thermal
anemometer in a constricted (3 inch high) access opening and applying appropriate
correction factor without adjusting cabinet airflow balance. The downflow nominal
set-point of 65 fpm was established 4 inches above the bottom of the sash with the I.V.
Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI 49-2014.
Certified canopy connection model is 4FT-CC.
[3] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer's operator's manual for appropriate correction factor) without
adjusting cabinet airflow balance. Downflow nominal set point of 65 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
49-2016.
[4] Inflow nominal set-point of 105 fpm was established with a direct airflow reading
instrument. A corresponding inflow nominal set-point of 105 fpm (corrected to local air
density) was confirmed using manufacturer's recommended alternate method with thermal
anemometer in a constricted (3 inch high) access opening and applying appropriate
correction factor without adjusting cabinet airflow balance. The downflow nominal
set-point of 65 fpm was established 4 inches above the bottom of the sash with the I.V.
Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI 49-2014.
Certified canopy connection model is 6FT-CC.
[5] Certified for use with power supplies of 115V/60/Hz, 230V/50Hz and 230V/60Hz.
|
|||||||||
Number of matching Manufacturers is 1
Number of matching Products is 8